摘要
本文使用BH1217型低本底α、β放射性测量仪,测定了生活饮用水中总α总β放射性体积活度,对影响测量结果的几个因素如标准源的测量、取水量的控制、水样蒸发、源制备的优化以及测量时间的控制等进行了分析,对有关放射性检测技术进行了探讨。
This article using BH1217 Low Background α, β Radioactivity Measuring Instrument, determination of total alpha total beta activity in drinking water, several factors that affect the measurement results, such as the selection of standard sources, control of water withdrawal, sulfation, optimization of source preparation, and control of measurement time were analyzed, discussed relevant radiation detection technology.
作者
龚锦瑜
Gong Jinyu(Guangdong Shunhua Environmental Technology Testing Co., Ltd., Foshan 528300, Chin)
出处
《广东化工》
CAS
2018年第11期244-245,共2页
Guangdong Chemical Industry
关键词
饮用水
总α放射性
总Β放射性
放射性检测技术
drinking water
total alpha radioactivity
total beta radioactivity
radioactive detection technology