2016年天津市部分区医用X射线摄影机质量控制状态检测结果与分析 被引量:6

Quality control status test and analysis of partial radiography unit in Tianjin
原文传递
导出
摘要 目的 通过对天津市6个行政区内172台医用X射线摄影机质量控制检测,了解天津市6个行政区医用X射线摄影机的性能指标状况。方法 根据医院级别和性质将医疗机构分为三级医院、二级医院、一级医院和民营医院,按照《医用常规X射线诊断设备影像质量控制检测规范》(WS 76—2011)标准对其中共性的9项指标进行检测,9项指标均合格的设备被判定为合格。采用SPSS16.0软件进行统计学分析。多组样本组间率的比较采用R×C列联表的χ2检验,多组样本间率的两两比较采用χ2分割法。结果 在172台设备中,合格的设备占69.2%;不合格指标主要是管电压指示的偏离和光野与照射野四边的偏离。三级医院设备合格率为82.4%,高于一级医院的51.9%和民营医院的54.1%,差异有统计学意义(χ2=9.272、9.599,均P〈0.01);三级医院国产设备占有率为7.35%,低于二级医院的27.5%、一级医院的92.6%和民营医院的78.4%,差异有统计学意义(χ2=7.907、64.184、54.478,均P〈0.01);二级医院国产设备占有率也低于一级医院和民营医院,差异有统计学意义(χ2=27.472、19.932,均P〈0.01);在用设备使用年限大于10年的,三级医院占16.2%,低于一级医院的51.9%,差异有统计学意义(χ2=12.933,P〈0.01)。结论 天津市6个行政区二级、三级医院医用X射线摄影机性能状态良好,一级和民营医院需进一步优化设备和加强质量控制管理。 Objective To understand tbhe performance indicators of radiography unit in six district of Tianjin. We tested and analyzed the testing results of 172 radiography unit in the light of quality control inspection and evaluation methods. Methods The medical institution was divide into tertiary hospital, secondary hospital, community hospital and private hospital. We tested the nine indexs of radiography unit with the Specifications for Testing of Image Quality Control in Medical X-ray Diagnostic Equipment(WS 76-2011). And only when the nine index were qualified, the radiography unit was judged to be qualified. All date were analysed by SPSS 16.0. Multi-sample rate was done by R×C Chi-square test, and multiple comparisons of rate was done by partition of Chi-square test method. Results The qualified rate of the 172 units were 69.2%. The unqualified indexs of radiography unit were the deviation of the pipe voltage indicator and the deviation of the light field from the four sides of the irradiation field. The qualified rate of state testing in tertiary hospital was 82.4%, which was higher than that 51.9% in community hospital and 54.1% in private hospital(χ2=9.272, 9.599, both P〈0.01), and these differences were all statistically significant. The occupancy rate of domestic radiography unit in tertiary hospital was 7.35%, which were lower than that 27.5% in secondary hospital, that 92.6% in community hospital and that 78.4%(χ2=7.907, 64.184, 54.478, all P〈0.01) in private hospital, and these differences were all statistically significant. And the occupancy rate of domestic radiography unit in secondary hospital were lower than that in community hospital and private hospital(χ2=27.472, 19.932, both P〈0.01), and these differences were all statistically significant. The rate of radiography unit used by tertiary hospitals over 10 years was 16.2%, which was lower than that 51.9%(χ2=12.933, P〈0.01) in community hospital, and this difference was statistically significant. Conclusion The performance status of radiography unit in tertiary hospital and secondary hospital of the six district of Tianjin are satisfactory, wherea community hospitals and private hospital need to optimize radiography unit and strengthen quality control management.
作者 魏超 高杰 张文艺 尹谌 翟贺争 武权 Wei Chao;Gao Jie;Zhang Wenyi;Yin Chen;Zhai Hezheng;Wu Quan(Tianjin Key Laboratory of Rdiation Medicine and Molecular Nuclear Medicine, Institute of Radiation Medicine Chinese Academy of Medical Sciences, Tianjin 300192, China)
出处 《国际放射医学核医学杂志》 2018年第3期257-260,共4页 International Journal of Radiation Medicine and Nuclear Medicine
基金 天津市自然科学基金重点项目(16JCZDJC36100)
关键词 X射线计算机摄影 质量控制 质量管理 X-ray computed radiograph Quality control Quality management
  • 相关文献

参考文献7

二级参考文献89

  • 1陈以水,王琦,李南翔,熊晓英,邓磊.江西省医用诊断X射线机性能现状调查分析[J].中华放射医学与防护杂志,2011,31(6). 被引量:8
  • 2国家技术监督局.JJG744-1997医用诊断X射线辐射源计量检定规程[M].北京:中国计量出版社,1998..
  • 3.RTI Electronics.Barracuda使用说明书[Z].北京:瑞典奥利科公司,2003..
  • 4中华人民共和国卫生部.放射诊疗管理规定,2006,01:24
  • 5中华人民共和国卫生部.WS76-2011医用常规X射线诊断设备影像质量控制检测规范[S].北京:中国标准出版社,2011-09-30.
  • 6吴毅,岳保荣,杜国生,等.医用常规Χ射线诊断设备影像质量控制检测规范[S].北京:中国标准出版社,2011.2-13.
  • 7张志兴,郑钧兴.医用Χ射线诊断卫生防护监测规范[S].北京:法律出版社,2002.1-4.
  • 8尉可道,岳保荣,程玉玺,等.计算机X射线摄影(CR)质量控制检测规范[S].北京:人民出版社,2007.1-4.
  • 9张志兴,郑钧兴,夏连季.医用Χ射线诊断卫生防护标准[S].北京:法律出版社,2002.1-6.
  • 10中华人民共和国卫生部.GBZ138-2002医用x射线诊断卫生防护监测规范[S].北京:中国标准出版社,2002.

共引文献23

同被引文献47

引证文献6

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部