期刊文献+

镍钛产品疲劳测试早期断裂原因分析

Cause Analysis of Early Fracture of Nitinol Device during Fatigue Test
下载PDF
导出
摘要 镍钛合金医疗器械产品,在疲劳测试过程中发生早期断裂,但该批原材料经检验是合格的。通过金相显微镜、扫描电镜及能谱等对产品断裂原因进行分析。结果表明,产品早期断裂是由于在承受最大应力的区域,局部存在夹杂物缺陷,致使产品在周期性疲劳应力作用下,产生早期疲劳断裂失效。 Early fracture of nitinol medical device occurred during the fatigue testing,but the raw material is qualified. The fracture of device was analyzed by means of metallographic microscope,SEM and EDS etc. The results show that the early fracture of the device is due to the local existence of the impurity defects in the areas subject to the maximum stress,resulting in fatigue failure of the device under cyclic fatigue stress.
作者 祁凤君 穆盈 廖经友 李晓玲 QI Feng-jun;MU Ying;LIAO Jing-you;LI Xiao-ling(Shenzhen Advanced Animal Study Service Center, Shenzhen Guangdong 518057, China)
出处 《热处理技术与装备》 2018年第3期53-57,共5页 Heat Treatment Technology and Equipment
关键词 镍钛合金 疲劳 硅胶管 夹杂物 nitinol fatigue silicone tube impurity
  • 相关文献

参考文献1

二级参考文献12

  • 1[1]T.Dnerig,A.Pelton,D.Stockel.An Overview of Nitinol Medical Applications.Materials Science and Engineering A273-275(1999)149-160
  • 2[2]The guidance for Industry and FDA Staff:Non-Clinical Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems.2005:9.Fatigue Analysis.10.Accelerated Durability Testing
  • 3[3]EN14299-2004(E):Non active surgical implants-particular requirements for cardiac and vascular implants-specific requirements for arterial stents 7.2.8 Fatigue testing
  • 4[4]ISO/CD 25539(2006):Cardiovascular implants-Endovascular devices-Part 2:Vascular stents 8.6.3.5.2 Stress/strain analyses,8.6.3.5.3 Fatigue safety factor determination,8.6.3.5.4 Fatigue durability testing
  • 5[5]M.Pease and B.Walsh.Nitinol constitutive law comparison for finite-element analysis.Proceedings of the first international conference on shape memory and superelastic technologies.2000,California:495-500
  • 6[6]W.J.Harrison and Z.C.Lin.The study of nitinol bending fatigue.Proceedings of the first international conference on shape memory and superelastic technologies.2000,California:391-396
  • 7[7]D.Tolomeo,S.Davidson and M.Santinoranont.Cyclic properties of superelastic nitinol:design implications.Proceedings of the first international conference on shape memory and superelastic technologies.2000,California:471-476
  • 8[8]C.Kulger,D.Maston and K.Perry.Non-zero mean fatigue test protocol for NiTi.Proceedings of the first international conference on shape memory and superelastic technologies.2000,California:409-417
  • 9[9]A.Pelton,X.-Y.Gong,T.W.Duerig "Fatigue Testing of Diamond-Shaped Specimens".Proceedings of SMST2003,Monterey,CA
  • 10[10]Richard Glenn,Jeff Lee.Accelerated Pulsitile Fatigue Testing of Ni-Ti Coronary Stents.Proceedings of the first international conference on shape memory and superelastic technologies.2000,California

共引文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部