期刊文献+

Octavius 4D系统稳定性验证分析 被引量:3

Stability analysis of Octavius 4D system
下载PDF
导出
摘要 目的:探讨Octavius 4D系统用于容积旋转调强放射治疗(VMAT)三维剂量验证的稳定性。方法:比较分析semiflex电离室和Octavius 729探测器阵列在6 MV、10 MV射束下对射野大小、剂量线性、剂量率线性和射野输出因子的响应。测量观察Octavius 4D系统旋转过程中角度仪示值与机架角的角度偏差。用3%/3 mm标准分析(10×10)cm^2旋转照射计划和VMAT计划机架归零与旋转照射的二维剂量分布;用3%/3 mm gamma分析标准评估VMAT计划。结果:预热剂量大于6 Gy是探测器稳定的必要条件。探测器阵列剂量响应是线性的,不同标称剂量率下剂量测量是稳定的。旋转照射过程中加速器机架角和Octavius 4D模体旋转角度误差在0.4°以内。6 MV和10 MV射束VMAT计划在分析标准为3%/3 mm时,三维剂量分布的平均通过率分别为96.03%和95.56%,满足临床计划剂量验证的标准。结论:Octavius4D模体联合Octavius 729探测器阵列是一套稳定性装置,用于治疗前验证VMAT计划是可靠的。 Objective To explore the stability of Octavius 4D system in the three-dimensional dose verification of volumetric modulated arc therapy(VMAT). Methods The responses of semiflex ionization chamber and Octavius 729 detector array to field size, dose linearity, dose rate linearity and output factors were assessed for 6 MV and 10 MV beams. When Octavius 4D system was running, the angular deviations between inclinometer and gantry were measured. With gantry in rotating state or fixed at 0°, the two-dimensional dose distributions of(10×10) cm^2 arc plans and VMAT plans were analyzed with 3%/3 mm criteria. VMAT plans were assessed by 3%/3 mm gamma analysis standard. Results A warm-up dose larger than 6 Gy was essential for detector stability. Dose response of detector array was linear, and dose-rate measurements were stable at different nominal dose rates.During rotating irradiation, the angular deviation of accelerator gantry angle and rotation angle of Octavius 4D phantom was found to be within 0.4°. When 3%/3 mm criteria was applied, the average gamma passing rate of three-dimensional dose distribution were 96.03% and 95.56% for VMAT plan with 6 MV or 10 MV beams. Conclusion Octavius 4D phantom combined with Octavius 729 detector array is a stable device and is reliable for pretreatment verification of VMAT.
作者 牛振洋 费振乐 陈志 NIU Zhenyang;FEI Zhenle;CHEN Zhi(University of Science and Technology of China, Hefei 230027, China;the 105 Hospital of the Chinese People's Liberation Army, Hefei 230031, China)
出处 《中国医学物理学杂志》 CSCD 2018年第6期628-632,共5页 Chinese Journal of Medical Physics
基金 国家重点研发计划(2017YFC0107504)
关键词 Octavius 4D系统 容积旋转调强放射治疗 探测器阵列 γ通过率 稳定性 Octavius 4D system volumetric modulated arc therapy detector array gamma passing rate stability
  • 相关文献

参考文献5

二级参考文献39

  • 1马金利,蒋国梁,傅小龙,廖源,吴开良,周莉钧.窄带野胶片检查在多叶准直器质量保证中作用探讨[J].中华放射肿瘤学杂志,2004,13(2):123-127. 被引量:10
  • 2戴建荣,胡逸民,张红志,关莹,张可,王闯.针对患者调强放射治疗计划的剂量学验证[J].中华放射肿瘤学杂志,2004,13(3):229-233. 被引量:65
  • 3孙朝阳,肖锋,石梅,魏丽春,陆军.Varian多叶准直器(MLC)常规测量方法及质量保证[J].中国医学物理学杂志,2005,22(6):697-698. 被引量:10
  • 4Clarke MF, Budgell GJ. Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle. Phys Med Biol,2008 ,53 :473-485.
  • 5Parent L,Seco J,Evans PM. Evaluation of two methods of predicting MLC leaf positions using EPID measurements. Med Phys, 2006,33:3174- 3182.
  • 6Childress NL, Dong L. Rapid radiographic film calibration for IMRT verification using automated MLC field. Med Phys, 2002,47, 4121-4130.
  • 7Bayouth JE, Wendt D, Morrill SM. MLC quality assurance techniques for IMRT application. Phys Med,2003,30:743-750.
  • 8Luo W, Li J, Price RA Jr, et al. Monte Carlo based IMRT dose verification using MLC log files and R/V outputs. Med Phys, 2006,33:2557-2564.
  • 9Zygmanski P, Kuang J, Jiang S, et al. Dependence of fluence errors in dynamic IMRT on leaf-positional errors varying with time and leaf number. Med Phys,2003 ,30 :2736-2749.
  • 10Woo MK, Nico A. Impact of multi-leaf collimator leaf positioning accuracy on intensity modulation radiation therapy quality assurance ion chamber measurements. Med Phys,2005,32:1440- 1445.

共引文献37

同被引文献16

引证文献3

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部