摘要
针对辐射光谱测温法测量误差问题,利用黑体炉搭建了标准辐射测温实验平台,选用200~1 100nm波段光谱仪对标准高温源进行辐射光谱测量。讨论了高温源辐射光谱特征,并基于辐射光谱测温法获得了温度测量值,该值与标准参考值的相对偏差小于4%,同时分析了测量重复性引起的标准不确定度分量,为辐射测温法应用提供参考。
In order to obtain the measurement error of radiation spectroscopy thermometry,the standard experimental platform was established by black body furnace.The spectrometer with 200~1 100 nm was selected to measure the radiation spectrums of standard high-temperature source.The spectral characteristics were discussed in detail.And the temperature results were determined based on spectroscopy thermometry.Compared with the standard value,the relative deviations are less than 4%.The uncertainty components of the repeated measurements were analyzed.It provides important reference for applications of radiation spectroscopy thermometry.
作者
ALLIOT Aymeric
徐秋丽
杨敬贤
郭浩然
杨斌
蔡小舒
ALLIOT Aymeric;XU Qiuli;YANG Jingxian;GUO Haoran;YANG Bin;CAI Xiaoshu(Shanghai Key Laboratory of Multiphase Flow and Heat Transfer in Power Engineering,University of Shanghai for Science and Technology,Shanghai 200093,China;Shanghai Space Propulsion Technology Research Institute,Shanghai 201109,China)
出处
《光学仪器》
2018年第3期8-11,共4页
Optical Instruments
基金
"十三五"装备预研国防科技重点实验室基金(61427010104)
上海航天科技创新基金(SAST2017-123)
关键词
辐射光谱
温度测量
光纤光谱仪
测量误差
radiation spectroscopy
temperature measurement
optical fiber spectrometer
measurement error