目的了解原某军区X射线诊断设备质量控制状态,提高放射防护水平。方法依据相关国家标准,对该军区各级医疗单位的X射线诊断设备进行质量控制状态检测。结果经质量控制状态检测,该军区X射线摄影设备合格率为84.7%,透视设备合格率为92.4%,总合格率为88.7%。未发现不同级别医疗单位间合格率的差异。结论该军区的Χ射线诊断设备质量存在着较多的问题,今后应进一步加强医疗单位,特别是低级别单位的X射线诊断设备质量控制检测和管理。
Objective To acquaint the basic status of quality control of X - ray diagnostic equipment in a military region so as to improve radiation protection level. Methods According to the related national criterions, quality control tests for medical X - ray diagnostic equipment were carried out in the medical institutions in a military region. Results The qualified rates of radiography and fluoroscopy equipment were 84.7% ,92.4% respectively. The total qualified rate was 88.7%. There was no statistical difference among different levels of medical institutions. Conclusion There existed some problems in quality control of X - ray diagnostic equipment in the military region. It is necessary to strengthen the quality control tests of X - ray diagnos- tic equipment, especially for the low level institutions.
作者
寇庆河
谢怀江
仝玉社
马力
KOU Qinghe;XIE Huaijiang;Huaijiang;TONG Yushe;MA Li(Sheng Yang Military Region Center for Disease Control and Prevention,Shenyang 110034 China;Epidemic Prevention Teams of the Support Department in Morthern Theater Command of the LAAF)
出处
《中国辐射卫生》
2018年第3期234-236,共3页
Chinese Journal of Radiological Health
关键词
X射线诊断设备
状态检测
质量控制
X-Ray Diagnostic Equipment
State Test