神经电极的水凝胶涂层及楔形角对组织损伤的影响 被引量:2

Effects of PVA-H Coating Thickness and Tip Angle on Tissue Injury
下载PDF
导出
摘要 目的研究水凝胶涂层的厚度以及电极楔形角对植入损伤的影响。方法基于植入损伤评估系统,进行模拟神经电极植入过程的实验,评估电极植入造成的组织损伤。用浸涂次数(分别为0、1、2、3)控制水凝胶涂层的厚度,选用30°、40°、50°、60°为电极楔形角的变量。以最大组织应变和最大植入力为组织损伤的衡量标准。结果水凝胶涂层越厚,电极植入损伤越大。楔形角越小,植入损伤也越小。同时,减小针尖的楔形角可以减小涂层对组织损伤的影响程度。3次浸涂时,楔形角为30°时,最大组织应变与最大植入力分别增加3.4%和3.8%,而楔形角为60°时,两者分别增加11.3%和18.1%。结论神经电极的水凝胶涂层将增大植入电极对生物组织的损伤,然而减小电极尖端楔形角的方法可以降低水凝胶涂层厚度对植入损伤的影响程度。 Objective To study the effects of PVA-H coating thickness and tip angle on the tissue injury caused by the implantation of neural electrodes. Methods Simulated implantation experiments were conducted based on a tissue injury evaluation system to evaluate the tissue injury caused by electrode implantation. The coating thicknesses were controlled by the number of dip coating times( 0,1,2,and 3),whereas the tip angles were set as 30°,40°,and 50°. The maximum tissue strain and insertion force were selected as the measurement of the tissue injury. Results A thicker hydrogel coating and larger tip angle would cause more serious tissue injury. Simultaneously,reducing the tip angle of the neural electrode could reduce the degree of the hydrogel coating effect on the tissue injury. When the tip angle was 30°,the maximum strain and the peak insertion force increased by 3. 4% and 3. 8%,respectively,whereas when the wedge angle was 60°,the maximum strain and maximum insertion force increased by 11. 3% and 18. 1%,respectively. Conclusions The hydrogel coating of the neural electrode increased the injury of biological tissues caused by the implantation of the neural electrode. However,the method of decreasing the tip angle of the electrode could reduce the degree of the negative effects of the hydrogel coating thickness on the implantation injury.
作者 唐嘉琪 张文光 尹雪乐 TANG Jiaqi;ZHANG Wenguang;YIN Xuele(School of Mechanical Engineering,Shanghai Jiao Tong University,Shanghai 200240,China)
出处 《医用生物力学》 EI CAS CSCD 北大核心 2018年第4期332-336,共5页 Journal of Medical Biomechanics
基金 国家自然科学基金项目(51675330)
关键词 神经电极 水凝胶 楔形角 组织应变 植入力 neural electrode hydrogel coating tip angle tissue strain insertion force
  • 相关文献

参考文献3

二级参考文献90

共引文献19

同被引文献3

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部