摘要
采用电感耦合等离子体质谱法测定废液晶屏中的金属含量。通过优化仪器工作条件,利用电感耦合等离子体质谱仪对废液晶屏的浸出液进行半定量分析,同时与全定量分析结果进行对比。结果表明:除Fe外,半定量与全定量测试的相对偏差为1.27%~18.95%,加标回收率为80.62%~118.52%,满足测试的要求,测定结果具有较高的参考价值。
The metal content in the waste liquid crystal display was determined by inductively coupled plasma mass spectrometry. Semi-quantitatively analysis of trace metals in the the waste liquid crystal display leaching solution was carried out by optimizing the instrument working conditions, and the results were compared with full quantitative analysis. The results indicate that the relative deviations of semi-quantitative and full-quantitative tests were 1.27 %~18.95 % except for Fe; the recoveries of standard addition were 80.62 %~118.52 %, which meet the requirements of the test, and the determination results have high reference value.
作者
陆静蓉
吴彤
朱炳龙
郭云飞
樊陆璐
金晓
周全法
Lu Jingrong;Wu Tong;Zhu Binglong;Guo Yunfei;Fan Lulu;Jin Xiao;Zhou Quanfa(School of Chemical and Environmental Engineering,Jiangsu University of Technology,Changzhou 213001;Jiangsu Key Laboratory of E-Waste Resource Recycling,Changzhou 213001;Changzhou Xiangyu Resource Recycling Technology Co.,Ltd.,Changzhou 213168,China)
出处
《广东化工》
CAS
2018年第18期39-41,共3页
Guangdong Chemical Industry
基金
江苏省前瞻性联合研究项目(BY2016030-25)
江苏省研究生实践创新计划项目(SJCX18-1002)
江苏省研究生创新计划项目(SJCX17_0774)
关键词
电感耦合等离子体质谱
废液晶屏
铟
分析
inductively coupled plasma mass spectrometry
waste LCD screen
indium