期刊文献+

基于高斯牛顿法的磁纳米温度快速反演算法 被引量:2

Fast inversion algorithm for magnetic nanoparticle temperature based on Gauss-Newton algorithm
原文传递
导出
摘要 根据磁纳米粒子磁化响应强度的奇次谐波信号,构建磁纳米粒子温度测量模型,并将温度模型的反演问题转化为非线性方程组的求解问题.采用一种基于改进型高斯牛顿法的温度快速反演算法,与传统高斯牛顿法进行对比,从温度反演角度研究不同算法对测温精度和时间分辨率的影响.仿真结果表明:该算法具有显著的高稳定性和快收敛性优点,能够达到磁纳米温度计在医学领域应用所需要的测温精度(0.1℃)、时间分辨率(1 s)的要求. According to the odd harmonic signal of magnetic nanoparticle magnetization response intensity,a temperature measurement model of magnetic nanoparticles was built,and the inversion problem of temperature model was transformed into nonlinear equations solving problem.A fast temperature inversion algorithm based on the improved Gauss-Newton algorithm was proposed,and compared with the traditional Gauss-Newton algorithm.The influence of different algorithms on the accuracy and time resolution of temperature measurement from the temperature inversion view was studied.The simulation results show that the improved Gauss-Newton algorithm has the obvious advantages of high stability and fast convergence,which can meet the requirements of temperature measurement accuracy of 0.1 ℃ and time resolution of 1 s for the application of magnetic nanoparticle thermometers in medical field.
作者 李娜娜 叶娜 苏日建 刘晓楠 Li Nana;Ye Na;Su Rijian;Liu Xiaonan(SchoolofComputerandCommunicationEngineering,ZhengzhouUniversityofLightlndustry,Zhengzhou450000,China;Dongfeng Nissan Passenger Vehicle Company Technical Center,Guangzhou 510800,China;School of Mechanical and Automotive Engineering,South China University of Technology,Guangzhou 510006,China)
出处 《华中科技大学学报(自然科学版)》 EI CAS CSCD 北大核心 2018年第10期105-109,132,共6页 Journal of Huazhong University of Science and Technology(Natural Science Edition)
基金 国家自然科学基金资助项目(61374014) 河南省科技攻关项目(172102210064) 中国博士后科学基金资助项目(2017M612660)
关键词 磁纳米温度计 温度反演 高斯牛顿算法 医学测温 癌症热疗 magnetic nanoparticle thermometer temperature inversion Gauss-Newton algnrithm medical temperature measure-ment cancer heat treatment
  • 相关文献

参考文献1

二级参考文献9

  • 1高深.人体皮下组织温度的微波测量方法[J].实用测试技术,1994,20(2):24-27. 被引量:2
  • 2Ogawa K et aI.A reconstruction algorithm from truncated projecrions.. IEER trans med imaging . 1995
  • 3XueH L,WeiY.A featreconstruction algorithm for diffractlon tomography. The Proc of the China-Japa-JointConlon Ultrason . 1988
  • 4Delphine G,PatrickC,AlexandreL,et a1.M Rmonitoring of laser-induced lesicms of the liverin vivoin alow-field openmagnet:temperatuermappingandlesionsizepredictino. Journal of Magnetic Resonance Imaging . 2001
  • 5DeIphine G,VahlaaE,EhnholmGJ,et a1.MR temperaturem Iementin-livIement in livertissueat 0.23T iwht asteady—slatefree precessionsequence. Mat-aResonMed . 2O02
  • 6Peter R,Christoph H,Carlo,M,et a1.MR-monitored L Tasapaliati reconcept in patients wiht hi gradegliomas:perliminary clinicalexperience. Journal of Magnetic Resonance Imaging . 1998
  • 7Kagayaki K,AnderwHC,KullervoH,et a1.Calibration of water protnochemmcial shift iwth temperatuer fornoninvasive temperature imagingduirng focused ultrasound surgery. Journal of Magnetic Resonance Imaging . 1998
  • 8Rudolf S,Peter WA,DietrichH,et al.Temperam monitoirng of inter-stitila thermaltissue coagulation using MR phaseimages. Journal of Magnetic Resonance Imaging . 1997
  • 9KentRT,RobertTycko.Measurement of sample temperatures underm-agic-angl espinning from thechemica lshift and spin latticerelaxationrate of 79BrinKBrpowder. Journa lo fMagneticResonance . 2009

共引文献5