防针刺伤塑料安瓿开口设计研究 被引量:2

New plastic ampoule opening design for use without needles: an evaluation study
原文传递
导出
摘要 目的本研究重新设计软包装安瓿的开口,使其与注射器的乳头相吻合,以有效避免针刺伤。同时通过实验研究,检测该开口设计的合理性。方法本研究通过3D打印技术制作新型软包装安瓿模型30支,并经环氧乙烷灭菌处理,并用注射器注入2 ml蒸馏水。后应用无针技术抽取出蒸馏水,计算抽取量与注入量的差异(流失量)。并注入1 ml蒸馏水于营养琼脂平板,在(35±1)℃培养48 h,计算菌落数。该实验重复执行2次。对照组采用60支2 m1无菌蒸馏水注射液,采用传统方法抽取液体,同上述方法计算量差(多余量)及菌落数。结果试验组无菌蒸馏水流失量中位数为0.06 m1.95%CI为0.05-0.07 ml;对照组无菌蒸馏水多余量中位数为0.06 ml,95%CI为0.06—0.07ml,2组比较差异无统计学意义(Z=-1.194,P=0.233)。试验组2次实验共培养出8个菌落数(CFU),对照组4个,2组比较差异无统计学意义(P=0.224)。结论该新型软包装开口有助于医护人员有效避免针刺伤。且该开口设计与注射器吻合度好,且不导致液体残留,无针抽药操作也不会增加污染率,提示该设计适用于临床实践。 Objective To produce a plastic ampoule with an opening that fits syringe adapters to allow solutions to be drawn up without needles (to avoid needle stick and sharps injuries) and to test the effectiveness of the new design in terms of residual fluid volume and risk of microorganism contamination. Methods Evaluation study based on laboratory research was adopted. For the experimental group, 30 ampoules based on the new design were produced by 3D printing. The ampoules were sterilized and filled with 2 ml sterile water. The sterile water was drawn up without the use of needles, and the wastage fluid (i.e., 2 ml minus the quantity drawn up) was calculated. A 1 ml aliquot of sterile water from each ampoule was dropped onto a nutrient agar plate, and the number of colony-forming units was assessed after 48 h. For each ampoule, the experiment was performed twice. Sixty 2 ml glass-packaged sterile water injections constituted the control group. The fluid was drawn up with a needle, and the superfluous fluid and number of colony-forming units were assessed, as in the experimental group. Results The mean wastage fluid was 0.06 ml, 95%CI was 0.05-0.07 ml in the experimental group and the superfluous fluid was 0.06 ml, 95%CI was 0.06-0.07 ml in the control group, there was no significant difference between the two groups (Z=-1.194, P=0.233). The number of colony-forming units was 8 in the experimental group and 4 in the control group, there was no significant difference between the two groups (P=0.224). Conclusions The newly designed plastic ampoule opening could help health workers to avoid needle stick and sharps injuries when drawing up solution. The wastage fluid and microorganism contamination levels met the required standards, indicating that the new design is suitable for clinical application.
作者 宋继红 闫丹萍 林美娟 郑朝婷 胡春洁 石莉娜 Song Jihong;Yan Danping;Lin Meijuan;Zheng Chaoting;Hu Chunjie;Shi Lina(School of Nursing,Fujian Medical University,Fuzhou 350001,China)
出处 《中国实用护理杂志》 2018年第31期2454-2457,共4页 Chinese Journal of Practical Nursing
基金 福建省社会科学规划项目(2012B019) 福建医科大学高等教育教学改革工程项目(J16019)
关键词 药品包装 设备改进 针刺伤 软包装安瓿 职业安全 Drug packaging Equipment design Needle stick injuries Plastic Equipment and supplies Occupational exposure
  • 相关文献

参考文献6

二级参考文献38

共引文献60

同被引文献30

引证文献2

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部