期刊文献+

流动注射法与纳氏试剂法测定水中氨氮之比较 被引量:5

Comparison of flow injection and nash reagent method for determination of ammonia nitrogen in water
下载PDF
导出
摘要 流动注射法是近年发展起来的一种分析方法,纳氏试剂比色法是国家的标准方法。本文对纳氏试剂光度法与流动注射分析法测定水中的氨氮进行对比实验,得出结论:流动注射法测定水中氨氮,检测范围0.01~10.0mg/L,分析速率为50个样品每小时,检出限为0.001mg/L,纳氏试剂法检测范围0.02~2.0mg/L,检出限为0.02mg/L。流动注射法操作简便,灵敏度、精准度符合分析工作要求,是较好的测定方法。 The fow injection method is an analytical method developed in recent years. Nessler’s reagent colorimetric method is the national standard method. This paper compares the ammonia nitrogen in water by spectrophotometry and fow injection analysis. It is concluded that the fow injection method is used to determine ammonia nitrogen in water. The detection range is 0.01~10.0mg/L. The rate is 50 samples per hour, the detection limit is 0.001 mg / L Nessler’s reagent detection range 0.02 ~ 2.0 mg / L, the detection limit is 0.02 mg / L. The fow injection method is easy to operate, and the sensitivity and precision meet the requirements of analytical work, which is a good method for determination.
作者 郑洁 Zheng Jie(Gaogang Environmental Monitoring Station,Taizhou City,Taizhou Jiangsu 225300,China)
出处 《环境与发展》 2018年第11期116-117,共2页 Environment & Development
关键词 流动注射法 纳氏试剂比色法 氨氮检测 Flow injection method Nessler reagent colorimetry Ammonia nitrogen detection
  • 相关文献

参考文献4

二级参考文献14

共引文献7

同被引文献34

引证文献5

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部