摘要
Two time-domain reflectometry (TDR) systems and a new impedance measuring instrument, Thetaprobe,which are based on determination of soil dielectric constant, were used to measure water content of clayeyred soil to eraluate the accuracy of these instruments. The results indicated that these instruments shouldbe carefUlly re-calibrated before being applied in clayey red soil. With a new calibration curve fed into one ofthe TDR systems tested, nase system, the measured data compared well with tho8e by standard oven-dryingmethod.
Two time-domain reflectometry (TDR) systems and a new impedance measuring instrument, Thetaprobe,which are based on determination of soil dielectric constant, were used to measure water content of clayeyred soil to eraluate the accuracy of these instruments. The results indicated that these instruments shouldbe carefUlly re-calibrated before being applied in clayey red soil. With a new calibration curve fed into one ofthe TDR systems tested, nase system, the measured data compared well with tho8e by standard oven-dryingmethod.
关键词
粘土状红壤
土壤粉
张力计
时间域反射测定
clayey red soil, dielectric constant, soil water content, time-domain refiectometry (TDR)