低电压扫描技术应用于CT血管成像对患者受检辐射剂量的影响

Effect of Low-Voltage Scanning Technique on Radiation Dose of CT Angiography
下载PDF
导出
摘要 目的探讨在CT血管成像CTA检查中应用低电压(100KV)扫描技术对患者受检辐射剂量减少的临床意义。方法选取2011年6月至2015年12月到本院进行CTA检查的300例患者,随机分成A组134例和B组166例。A组为低电压组(100 kV),B组为常规电压组(120 kV)。管电流均为450 mAs。比较两组的CTA对患者辐射剂量及血管图像质量。结果两组图像质量均能满足诊断要求。A组辐射剂量明显低于B组(P <0.05)。结论 CTA成像采用低电压在保证图像质量前提下可以有效减低患者辐射剂量。 Objective To investigate the clinical significance of low voltage(100 KV)scanning technique in CT angiography for reducing radiation dose. Methods TA total of 300 patients admitted to our hospital for CTA examination from June 2011 to December 2015 were randomly divided into group A(134 cases) and group B(166 cases). Group A is the low voltage group(100 kV), and group B is the normal voltage group(120 kV).and tube current was 450 m As. The radiation dose and vascular image quality of CTA were compared between the two groups. Results Both groups of images could meet the diagnosis requirements. The radiation dose of group A was significantly lower than that of group B(P<0.05). Conclusion CTA imaging with low voltage can effectively reduce the radiation dose of patients under the premise of ensuring image quality.
出处 《大医生》 2018年第10期113-114,共2页 Doctor
关键词 计算机体层摄影 低电压 CT血管成像 辐射剂量 computer body photography low voltage CT angiography radiation dose
  • 相关文献

参考文献4

二级参考文献64

  • 1梁国庆,赵玉娟,姜铁民,陈少伯,赵季红,李玉明.血管内超声虚拟组织学对急性冠脉综合征患者冠状动脉斑块诊断的临床研究[J].中国中西医结合急救杂志,2010,17(2):108-110. 被引量:3
  • 2Schueller-Weidekamm C,Schaefer-Prokop CM,Weber M,et al.CT angiography of pulmonary arteries to detect pulmonary embolism:improvement of vascular enhancement with low kilovoltage settings.Radiology,2006,241:899-907.
  • 3Heyer CM,Mohr PS,Lemburg SP,et al.Image quality and radiation exposure at pulmonary CT angiography with 100-or 120 kVp protocol:prospective randomized study.Radiology,2007,245:577-583.
  • 4Takahashi M,Maguire WM,Ashtari M,et al.Low-dose spiral computed tomography of the thorax:comparison with the standarddose technique.Invest Rediol,1998,33:68-73.
  • 5Lucaya J,Piqueras J,Garcia-Pena P,et al.Low-dose high resolution CT of the chest in children and young adults:dose,cooperation,artifact incidence,and image quality.AIR,2000,175:985-992.
  • 6Diederich S,Lenzen H,Puskas Z,et al.Low dose computerized tomography of the thorax:experimental and clinical studies.Radiology,1996,36:475-482.
  • 7Huda W,Scalzetti EM,Levin G.Technique factors and image quality as functions of patient weight at abdominal CT.Radiology,2000,217:430-435.
  • 8Primak AN,McCollough CH,Bruesewitz MR,et al.Relationship between noise,dose,and pitch in cardiac multi-detector row CT.Radiographics,2006,26:1785-1794.
  • 9Ertl-Wagner BB,Hoffmann RT,Bruning R,et al.Multi-detector row CT angiography of brain at various kilovoltage settings.Radiology,2004,231:528-535.
  • 10Naidich DP,Marshall CH,Gribbin C,et al.Low-dose CT of the lungs:preliminary observations.Radiology,1990,175:729-731.

共引文献103

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部