期刊文献+

SIGNAL SAMPLING DEPTH-INDUCED SKEW EFFECT IN SIMS DEPTH PROFILING

SIGNAL SAMPLING DEPTH-INDUCED SKEW EFFECT IN SIMS DEPTH PROFILING
下载PDF
导出
摘要 The effect of signal sampling depth in SIMS on profile was not reported We have found that the maximum sampling depth in SIMS is considerably bigger than the mean penetration range of primary ions and can skew the profile of secondary ion counts for implanted samples on the more deeper direction from the surface. The effect of maximum sampling depth is true not only for implanted samples but also for the samples in the middle of which there is an other impurity-rich layer. The action principle of signal sampling depth effect and the method of decreasing the error produced by the effect are discussed in this The effect of signal sampling depth in SIMS on profile was not reported We have found that the maximum sampling depth in SIMS is considerably bigger than the mean penetration range of primary ions and can skew the profile of secondary ion counts for implanted samples on the more deeper direction from the surface. The effect of maximum sampling depth is true not only for implanted samples but also for the samples in the middle of which there is an other impurity-rich layer. The action principle of signal sampling depth effect and the method of decreasing the error produced by the effect are discussed in this paper
出处 《真空科学与技术学报》 EI CAS CSCD 1992年第Z1期171-174,共4页 Chinese Journal of Vacuum Science and Technology
基金 This work is supported by National Natural Science Fundation.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部