CT三代扫描旋转中心偏离的卷积反投影算法 被引量:10

Convolution Back-projection Reconstruction forⅢ CT Scanning with a Shift Center of Rotation
下载PDF
导出
摘要 讨论了卷积反投影算法实现CT图像重建时 ,被测工件旋转中心偏离理想重建中心的工程问题。基于理想卷积反投影重建算法理论 ,推导出相应校正偏离的卷积反投影算法 ,并引入单象素工件模型估计偏离参数 ;根据偏离幅度给出了理想重建算法适应的范围 ,仿真结果证明了校正算法的有效 ,利用校正算法消除了由旋转中心偏离造成的伪影 。 The problem of displacement of center of rotation is discussed in CT image reconstruction by convolution back projection. Basing on ideal convolution back projection, a convolution back projection with a shift center of rotation has been derived for correcting images that acquired with a divergence center of rotation using the fan beam geometry with an angle equaled detector. In order to solve parameter of divergence axis of rotation, a single pixel phantom has been introduced. Software simulations are performed to certify correcting method's feasibility, as a result, eliminate the tail off artifacts and improve the resolution of CT image.
出处 《重庆大学学报(自然科学版)》 EI CAS CSCD 北大核心 2002年第9期55-58,共4页 Journal of Chongqing University
关键词 CT 三代扫描 辐射源中心 旋转中心 卷积反投影 Radon逆变换 the third scanning method center line of reconstruction center of radiation Center of rotation convolution back projection, radon transformation