摘要 在自行设计研制的小口径针织机上 ,从牵拉张力、弯纱深度、给丝张力等方面探讨了针织医用金属内支架的编织工艺对支架形态的影响规律 ,揭示了编织规律和不同编织工艺条件下的较佳值。 This paper mainly discusses the influences in knitting process such as the taken-down tension, sinking depth and feeding tension. These three factors are separately changed to understand the relationship between the knitting process and the shape of stent. The results show that the knitting technology significantly influences the shape of stent and the optimal knitting parameters were obtained by using the regression method.
出处 《东华大学学报(自然科学版)》 CAS CSCD 北大核心 2002年第5期30-33,共4页 Journal of Donghua University(Natural Science)
关键词 针织 医用 金属内支架 编织工艺 记忆合金 knit, medical, stent, shape memory alloy
  • 相关文献

参考文献3

二级参考文献23

  • 1[1]Nishijima E, Muraji T, Yoshimura M. A shape memory airway stent for tracheobromalacia in children: An experimental and clinical study [J] . Pediatri Surg, 1997,32(1) :50 - 53
  • 2[2]Cushing WJ, Magovern GJ, Olearchyk AS.Internal mammary artery graft: retrospective report with 17 years' survival [J] .Thorac Cardiovas Surg, 1986,92(5) :963 - 964
  • 3[3]Sigwart 131H. Coronary endoprostheses, 1990,15(5):319-328
  • 4[4]Dai K, Chu Y. Studies and applications of NiTi shape memory alloys in the medical field in China [ J] . Biomed Mater Engl, 1996,6(4) :233 - 240
  • 5[5]Trepanier C, LeungTK, Tabrizian M. Preliminary investigation of the effects of surface treatments on biological response to shape mem ory NiTi stents [J] . Journal Biomedical Material Research, 1999,48(2):165- 171
  • 6[6]Gotman I.Characteristics of metals used in implants [J] .Endouro, 1997,11(6) :383 - 389
  • 7[7]Teitelhaum GP, Bradley WG Jr, Klein BD.MR imaging artifacts, ferromagnetism, and magnetic torque of intravascular filters, stents,and coils [J] . Radiol, 1988,166(3) :657 - 664
  • 8[8]Engellau L, Olsrud J, Brockstedt S, et al. MR evaluation in vivo and in vivo of a covered stent- graft for abdominal aortic aneurysms:Ferromagnetism, heating, artifacts, and velocity mapping [J] .Magne Reson Imag,2000,12(1):112- 121
  • 9[9]Gimenez AA, Riambau V, Serra BE,et al. Metal- induced generalized pruriginous dermatitis and endovascular surgery [J] .Contact Dermatitis, 2000,43(1) :35 - 40
  • 10[10]Holmes SA, Miller PD, Crocker PR, et al. Encrustation of intraprostatic stents - a comparative study [J] . Br J Urol, 1992,69(4) :383- 387

共引文献7

同被引文献32

引证文献4

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部