期刊文献+

100L/min浮游菌采样器流量校准工况 被引量:5

Flow rate calibration condition for 100 L/min airborne microbe samplers
下载PDF
导出
摘要 通过在采样工况条件下获得100 L/min浮游菌采样器的工况阻力实验,设计了合理的100 L/min浮游菌采样器的流量校准工况条件和模拟工况装置,使浮游菌采样器在使用模拟工况装置状态下获得的校准流量更接近于实际采样状态下的流量,为浮游菌采样器的采样流量校准提供了一定的实验依据。 This paper obtained the working condition resistance of 100L/min airborne microbe samplers under the sampling condition byexperiments, and designed a reasonable flow rate calibration condition anda working condition simulator for 100 L/min airborne microbe samplers.By using the simulator during calibration the calibrated flow rate of theairborne microbe sampler approached better the real flow rate under actualsampling condition. This paper provided some experimental basis forcalibrating the sampling flow rate of airborne microbe sampler.
作者 陈启悦 刘悦 施禅臻 蔡泽仁 张爱民 Chen Qiyue;Liu Yue;Shi Chanzheng;Cai Zeren;Zhang Aimin(Shanghai Institute of Measurement and Testing Technology;Tetra Park (Kunshan) Co., Ltd.))
出处 《上海计量测试》 2016年第4期46-48,共3页 Shanghai Measurement and Testing
关键词 100L/min 浮游菌采样器 采样流量 校准工况 100 L/min airborne microbe sampler sampling flow rate calibration condition
  • 相关文献

参考文献3

二级参考文献4

共引文献7

同被引文献11

引证文献5

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部