期刊文献+

气溶胶稀释器的稀释比校准方法 被引量:2

Calibration method of aerosol dilution device
下载PDF
导出
摘要 选用气溶胶光度计法、光学粒子计数器法和流量测定法对气溶胶稀释器的稀释比进行校准。描述了三种校准方法的校准过程和校准数据,并进行分析、比较,给出了三种校准方法的优劣和适用性,以及在气溶胶稀释器稀释比校准工作的注意事项。 The dilution ratio of aerosol dilution device was respectivelycalibrated with aerosol photometer method, optical particle countermethod and flow rate determining method in the paper. The calibrationprocess and data of three calibration methods were described, analyzedand compared. Advantages, disadvantagesare and the applicability ofthree calibration methods as well as matters needing attention on aerosoldiluter dilution ratio calibration were given.
作者 张爱亮 Zhang Ailiang(Shanghai Institute of Measurement and Testing Technology)
出处 《上海计量测试》 2016年第4期52-54,58,共4页 Shanghai Measurement and Testing
关键词 气溶胶稀释器 稀释比 校准 aerosol dilution device dilution ratio calibration
  • 相关文献

参考文献2

二级参考文献4

  • 1GB 13554—92《高效空气过滤器》,1992.
  • 2European Committee For Standardization .EN 1822-5 High efficiency particulate air filter (HEPA and ULPA)-part 5:Determining the efficiency of filter element 2000.
  • 3Standard Test Method For Air Cleaning Performance Of A High-Efficiency Particulate Air Filter System(ASTM F1471-93), 1993.
  • 4GB 6165-85《高效过滤器性能试验方法透过率和阻力》.1985.

共引文献1

同被引文献12

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部