摘要
本文利用青霉素酶与青霉素发生水解反应生成青霉噻唑酸,该酸电离出H+,促使溶液中的氧化苏木精发生还原反应,产生电流信号。以玻碳电极为工作电极,建立一种检测青霉素的电化学方法。在最佳的实验条件下,青霉素浓度在10-9~10-5 mol·L^(-1)范围内与其相对电流强度呈良好的线性关系(r2=0.9973),最低检出限(S/N=3)为1.661×10-8 mol·L^(-1)。连续平行测定1×10-6 mol·L^(-1)的青霉素溶液3次,相对电流强度的RSD为2.50%,表明该方法具有良好的稳定性。将该方法用于牛奶样品的实际测定,回收率为99.70%~102.69%,结果较为理想。
The penicillinase catalyzed the hydrolysis of penicillin to penicilloic acid, which H+ was liberated. H+ promoted the reduction reaction of hematein, thereby generated a current signal. Based on the principle, a new electrochemical method was established for the determination of penicillin at the glass carbon electrode. Under the optimum conditions, the GCE/hematein/penicillinase exhibited excellent performance for penicillin in PBS with a wide range of 10-9~10-5 mol/L and the detect limit was 1.661×10-8 mol/L(S/N=3). The RSD for 1×10-6 mol/L penicillin was 2.50%, which indicated that the method had good stability. The detection of penicillin concentration in real sample (milk) had acceptable accuracy with the assay system, which the recovery was 99.70%~102.69%.
作者
杜寒春
冯德芬
叶赛芳
黄泽军
艾晨昊
罗燕妮
谭学才
DU Hanchun;FENG Defen;YE Saifang;HUANG Zejun;AI Chenhao;LUO Yanni;TAN Xuecai(Guangxi Zhuang Autonomous Region Center for analysis and Test Research,Nanning 530022, China;School of Chemistry and Chemical Engineering, Guangxi University for Nationalities, Key Laboratory of Guangxi Colleges and Universities for Food Safety and Pharmaceutical Analytical Chemistry, Nanning 530008, China)
出处
《化工技术与开发》
CAS
2016年第12期30-34,48,共6页
Technology & Development of Chemical Industry
基金
国家自然科学基金项目(21365004)
广西直属公益性科研院所基本科研业务费专项项目(2013ACZ04)