CT图像均匀性数据的质量管理 被引量:9

Quality Control in Uniformity of CT Image Data
下载PDF
导出
摘要 目的将统计学方法及统计过程控制,应用于CT图像均匀性数据的质量管理中,以提高CT图像质量,更好地为肿瘤放疗患者提供更加清晰的图像。方法根据JJG 1026-2007国家计量检定规程中对CT均匀性检测方法及标准,检测我院肿瘤放疗模拟定位CT机的均匀性,并利用统计过程的控制策略处理检测数据并绘制控制图。结果根据控制图得出,我院肿瘤放疗模拟定位CT机在刚开机预热后至运行到需要散热的过程中符合正态分布曲线,CT图像均匀。在同一k V值下m A值在200 m A左右时CT图像均匀性较为稳定;在同一m A值下,k V值在130 k V时CT图像的均匀性较为稳定。结论建议我院肿瘤放疗科的CT模拟定位机的操作技师在开机预热后可以空机运行1次;尽可能选择机器运行参数在200 m A、130 k V左右,使CT图像的均匀性保持在一个很好的状态中;保持CT机房合适的环境温度。 Objective To apply statistical method and statistical process control to quality management on uniformity of CT image data, which was aimed at improving the quality of CT images and better providing more clear CT images for patients of tumor radiotherapy (RT). Methods The uniformity of simulating location CT device for tumor RT in our hospital was detected, and the adopted detectionmethod was referred to verification method for uniformity of CT images, which was described in Verification Regulation of Medical Diagnostic X-ray Radiation Source for Spiral CT (JJG 1026-2007 ofNational Metrology Verification Regulation of PRC). Then the data gained from detection above weretested and disposed through the control strategy of statistical process so as to implement a control chartdrawing. Results The control chart showed that uniformity of CT image data, which were obtainedon the operating machine from the time after warm-up to the time before heat dissipation, accordedwith normal distribution curve. Besides, under the fixed tube voltage (same kV value), uniformity ofCT images was relatively stable when tube current was about 200 mA. In the other hand, uniformity ofCT images was relatively stable if tube voltage was around 130 kV in view of the same tube current.Conclusion The technician, who operated simulating location CT device for tumor RT in our hospital,was advised firstly to do no-load operation once after the warm-up, secondly to maintain operation parameters of CT device at 200 mA, 130 kV more or less, keeping the CT image uniformity at a good state. In addition, the ambient temperature in CT device room should be set suitable.
作者 卜羽 高关心 BU Yu;GAO Guan-xin(Medical Engineering Section, Department of Medial Engineering, Inner Mongolia People’s Hospital, Huhhot Inner Mongolia 010010, China)
出处 《中国医疗设备》 2017年第2期47-49,共3页 China Medical Devices
关键词 CT模拟定位机 图像均匀性 肿瘤放疗 统计学方法 X射线辐射源 simulating location CT device image uniformity tumor radiotherapy statistical method X-ray radiation source
  • 相关文献

参考文献7

二级参考文献28

  • 1李新华,李强,张海宁,崔文国,李荣环,张振.医用CT的质量控制和保证[J].医疗卫生装备,2004,25(7):39-39. 被引量:16
  • 2杨裕华.X线CT影像技术研究进展[J].医疗设备信息,2004,19(9):25-28. 被引量:12
  • 3周代全,牟玮,巫北海,丁仕义.CT透视技术的发展和放射线剂量分析[J].第三军医大学学报,2004,26(21):1977-1979. 被引量:2
  • 4[2][美]Jiang Hsieh.计算机断层成像技术[M].北京:科学出版社,2005.
  • 5[5]GB/T11755.1-1989,医用诊断x射线机管电压测试方法[S].
  • 6[7]GB/T 17589-1998,X射线计算机断层摄影设备影响质量保证检测规范[S].
  • 7[8]于建明.医学影像技术学[M].北京:科学技术出版社,2004:215-221.
  • 8[9]Specification and Acceptance testing of Computed Tomography Scanners[R].AAPM,KEPORT 1993,39:26-39.
  • 9蒋国梁.现代肿瘤放射治疗学[M].上海:上海科技出版社,2002.45-69.
  • 10Goodenough D,Atkins F,Dyer S.Automated techniques for quality assurance of radiological image modalities.SPIE.Image Capture,Formatting,and Display,1991,1444:87-99.

共引文献50

同被引文献70

引证文献9

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部