数字化晨检仪的研制 被引量:4

Development of the Digital Morning Check Device
下载PDF
导出
摘要 目的研制用于直线加速器输出量晨检的数字化晨检仪。方法采用单个平板电离室作为探测器,与PIC16F877单片机处理器等构成数字化检测系统。按照JJG912-2010对治疗型电离室剂量检定的要求,在直线加速器下进行输出剂量测量,对系统的剂量学特性进行测试。结果在常规的跳数检测范围内(50~400 MU),晨检仪的测量剂量线性较好,线性拟合的相关系数均近似为1,最大偏差为0.4%。单次测量中,其测量重复性的最大偏差好于0.3%。对于不同剂量率的响应,其最大偏差小于0.3%。在半年检测过程中,其稳定性好于1%。结论研制晨检仪的剂量特性符合JJG912-2010《治疗水平电离室剂量计》的要求,可用于直线加速器输出量的晨检。 Objective To develop a digital device for measuring the dose output of medical accelerator in the morning check. Methods It was constructed by a single tablet ionization chamber and PIC16F877 single chip processor. The measurement of output dose under the medical accelerator was performed according to the document of JJG912-2010, and the dosimetry characteristics of the medical accelerator was tested by morning check device. Results In the range of 50~400 MU, the measured dose was linear.The correlation coefficients of linear fitting were approximately 1.00 and the maximum deviation was smaller than 0.4%. In a single measurement, the maximum deviation of measurement repeatability was better than 0.3%. The maximum deviation was less than 0.3% for different dose rates. In the long-term (half-year) testing process, the reproducibility of device was better than 1%. Conclusion Dosimetry characteristics of the device can meet the requirements of JJG912-2010 Verification Regulation of Ion Chamber Dosimeters Used in Radiotherapy. The research and development of the device can be used to measure the relative dose of the Linac in daily tests.
作者 张咏波 陈立新 ZHANG Yong-bo;CHEN Li-xin(Department of Equipment;Department of Radiotherapy, State Key Laboratory of Oncology in Southern China (Cancer Center of Sun Yat-sen University),Guangzhou Guangdong 510060, China)
出处 《中国医疗设备》 2017年第4期42-45,共4页 China Medical Devices
基金 广东省教育部产学研重点项目(2012B091000144)
关键词 放射治疗 直线加速器 电离室 剂量特性 单片机处理器 晨检仪 radiation therapy medical accelerator ionization chamber dosimetry characteristics single chip processor morning check device
  • 相关文献

参考文献10

二级参考文献92

  • 1黄晓延,黄劭敏,张黎,邓小武,钟宁山,陈立新.三维治疗计划系统的剂量学验证[J].中华放射肿瘤学杂志,2006,15(6):496-500. 被引量:26
  • 2ATTIX F H. Radiation dosimetry: Vol. 1[M]. New York: Pergamon Press, 1966.
  • 3阿蒂克斯,罗奇.辐射剂量学[M].施学明,陈常茂,于耀明,等译.北京:原子能出版社,1981.
  • 4International Electrotechnical Commission. Medical electrical equipment-Dosimeters with ionization chambers as used in radiotherapy[R]. Geneva: International Electrotechnical Commission, 1997.
  • 5BOAG J W. Ionization chamber[M]// Radiation dosimetry. New York : Press Academic, 1968.
  • 6KEMP L A W. Second 2 standard ionization chamber design[J]. Phys Med Biol, 1970, 15: 557-558.
  • 7AIRD E G A, FARMER F F. The design of a thimble chamber for the farmer dosimeter[J]. Phys Med Biol, 1972, 17: 169-174.
  • 8International Eleetrotechnical Commission. Dosi meters with ionization chambers as used in radiotherapy[R]. Geneva: International Electrotech nical Commission, 1994.
  • 9ICRU. Beam production, beam delivery and measurement of absorbed dose international commission on radiation units and measurement 7910 woodmont avenue bethesda, ICRU Report 59 [R]. Maryland, USA: ICRU, 1998.
  • 10IAEA. Absorbed dose determination in external beams radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water, IAEA Technical Reports Series No. 398[R]. Vienna: IAEA, 2000.

共引文献50

同被引文献27

引证文献4

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部