期刊文献+

多层螺旋CT图像常见伪影的影响因素与解决方案 被引量:30

Discussion on influence factors of common artifact of MSCT images
下载PDF
导出
摘要 目的:探讨多层螺旋CT成像过程中常见伪影的影响因素与解决方案,提高图像质量,获取优质的CT影像。方法:选取139例患者64层螺旋CT扫描后的影像资料,其中头部伪影38例、胸部伪影53例、腹部伪影29例以及盆腔伪影19例,对CT图像伪影的形态和成因逐一进行分析。结果:头部伪影包括:射线束硬化伪影24例、探测器故障伪影1例、斑马样伪影2例、运动伪影1例、金属伪影10例;胸部伪影包括:射线束硬化伪影8例、部分容积效应伪影9例、光子不足伪影9例、欠采样伪影1例、散射伪影3例、运动伪影21例、金属伪影1例、投影数据不完全伪影1例;腹部伪影包括:部分容积效应伪影10例、散射伪影5例、运动伪影4例、金属伪影10例;盆腔伪影包括:光子不足伪影5例、金属伪影14例。分析产生伪影的原因,主要为设备固有因素与检查技术因素两类;提出相应解决方案。结论:多层螺旋CT图像的伪影与设备固有及成像检查技术的诸多因素有关,优化扫描技术参数,可有效消除、抑制或避免图像伪影,获得优质图像。 Objective:To discuss the influence factors and solution of common artifact in the imaging process of multi-slice computed tomography(MSCT),so as to improve the image quality and obtain better quality for image document.Methods:139patients once were scanned by using PHILIPS64-MSCT were enrolled and their documents were analyzed,and these cases included38cases with head artifact,53cases with chest artifact,29cases with abdominal artifact and19cases with pelvic artifact.To analyze the morphology and cause of every CT image artifact.Results:In38cases with head artifact,24cases were caused by hardening of bundle of rays,1case was caused by fault of detector,2cases were caused by interpolating algorithm,1case was caused by motion,10cases were caused by metal material.In53cases with chest artifact,8cases were caused by hardening of bundle of rays,9cases were caused by partial volume effect,9cases were causedby insufficient photon,1case was caused by under-sampling,3cases were caused by scattering,21cases was caused by motion,1case was caused by metal material,and1case was caused by data incomplete of projection.In29cases with abdomen artifact,10cases were caused by partial volume effect,5cases were caused by scattering,4cases was caused by motion,and10cases were caused by metal material.In19cases with pelvic artifact,5cases were caused bym insufficient photon,and14cases was caused by metal material.Conclusion:The artifact of MSCT image are related to many factors,the optimizing scan parameters can effectively eliminate,suppress or avoid image artifact,and obtain high quality image.
作者 钱根年 陈自谦 付丽媛 赵春雷 陈蕾蕾 QIAN Gen-nian;CHEN Zi-qian;FU Li-yuan
出处 《中国医学装备》 2017年第5期18-22,共5页 China Medical Equipment
关键词 多层螺旋CT 伪影 图像处理 图像质量 原因分析 质量控制 Multi-slice spiral CT Artifact Image processing Image quality Cause Quality control
  • 相关文献

参考文献11

二级参考文献94

共引文献80

同被引文献276

引证文献30

二级引证文献100

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部