期刊文献+

ATP荧光监测技术在眼科负压吸引环清洗质量监测中的应用 被引量:3

Application of ATP Fluorescence Monitoring Technique in Quality Inspection of Ophthalmic Negative Pressure Suction Ring Cleaning
下载PDF
导出
摘要 目的:应用三磷酸腺苷(ATP)生物萤光法检测眼科负压环的清洗质量。方法:选取本院眼视光中心眼科负压吸引环120件,随机分为两组,分别为A组(预浸泡2min+酶清洗剂泡3min+超声3min)、B组(刷洗1min+酶清洗剂泡3min+超声3min)对两组的清洗效果进行比较。结果:B组的效果显著优于A组(P<0.05)。结论:ATP生物萤光法可动态快速监测眼科负压环清洗前后的细菌残留量,能作为有效监测眼科负压吸引环清洗质量的评价方法;刷洗+多酶清洗剂浸泡+超声清洗是负压吸引环的重要清洗环境。
作者 廖志利 李秋瑜 易红 LIAO Zhi-li;LI Qiu-yu;YI Hong(Department of Surgery, the Fourth Affiliated Hospital of Guangxi Medical University,Guangxi Liuzhou 545005)
出处 《中国医疗器械信息》 2017年第15期129-130,共2页 China Medical Device Information
基金 广西壮族自治区卫生和计划生育委员会2015年计划课题(合同号:Z2015670)
  • 相关文献

参考文献3

二级参考文献51

  • 1Haddad D, Worst J G. The maintenance and repair of ophthal- mic surgical instruments: training at the eye clinic[J]. Com- munity Eye Health, 2002,15(44) : 60-61.
  • 2Stevens S, Mason I. Care of ophthalmic surgical instruments [J]. Community Eye Health,2011,24(76) :36-37.
  • 3Ayaki M, Shimada K, Yaguehi S, et al. Corneal and conjunc- tival toxicity of disinfectants--assessing safety for use with ophthalmic surgical instruments [J]. Regul Toxicol Pharma- col,2007,48(3) 292-295.
  • 4Davidson CA, Griffith C J, Peters AC, et al. Evaluation of two methods for monitoring surface cleanliness - ATP bioluminescence and traditional hygiene swabbing [J]. Luminescence, 1999, 14 (1) :33 -38.
  • 5Alfa M J, Fatima I, Olson N. The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channels [ J ]. Am J Infect Control,2013, 41 (3) :249 -253.
  • 6Hansen D, Benner D, Hilgenhoner M, et al. ATP measurement as method to monitor the quality of reprocessing flexible endo- scopes [J]. Ger Med Sci J, 2004,(2) :04.
  • 7Obee PC, Griffith C J, Cooper RA, et al. Real - time monitoring in managing the decontamination of flexible gastrointestinal endo- scopes [ J ]. Am J Infect Control,2005,33 (4) :202 - 206.
  • 8Fushimi R, Takashina M, Yoshikawa H, et al. Comparison of a- denosine triphosphate, microbiological load, and residual protein as indicators for assessing the cleanliness of flexible gastrointesti- nal endoscopes [ J]. Am J Infect Control, 2013,41 ( 2 ) : 161 - 164.
  • 9Alfa MJ, Olson N, Murray BL. Comparison of clinically relevant benchmarks and channel sampling methods used to assess manual cleaning compliance for flexible gastrointestinal endoscopes [ J]. Am J Infect Control ,2014,42 ( 1 ) : 1 - 5.
  • 10Marena C, Lodola L, Zecca M, et al. Assessment of handwash- ing practices with chemical and microbiologic methods: prelimi- nary results from a prospective crossover study [ J ]. Am J Infect Control, 2002,30 (6) :334 - 340.

共引文献53

同被引文献40

引证文献3

二级引证文献4

相关作者