摘要
目的:探讨气相色谱法测定空气中苯的影响因素,分析可能存在的误差来源。方法:根据GB50325中要求进行采样和分析。结果:试样中苯浓度为1.5μg/m L对苯的稳定性影响显著,其他试样的批内、批间变异分析不显著。结论:低浓度苯在二硫化碳介质中存在不稳定性因素,随放置时间的延长苯有变异的趋势。
Objective:To study the influence factors of benzene in air by gas chromatography,analysis of possible sources of error.Method:Sampling and analysis are required in accordance with GB50325.Result:The concentration of benzene in the sample is1.5g/mL,which has remarkable influence on the stability of benzene.There were no significant differences between the batch and batch analysis of other samples.Conclusion:Unstable factors exist in low concentration benzene in carbon disulfide medium.With the extension of time,benzene has a trend of variation.
作者
王小江
刘芳
辜双娟
Wang Xiaojiang;Liu Fang;Gu Shuangjuan(Guizhou Guicai Environmental Monitoring Co., Ltd.,Guiyang Guizhou 550018)
出处
《河南科技》
2017年第15期134-135,共2页
Henan Science and Technology
关键词
苯
影响因素
空气
benzene
influencing factors
air