绝对剂量的偏移对MatriXX剂量验证结果的影响 被引量:6

Influence of Deviation in Absolute Dose on Verification Results of MatriXX
下载PDF
导出
摘要 目的探讨加速器绝对剂量的偏移对调强计划射野的Matri XX剂量结果影响。方法选取24例宫颈癌患者调强计划中机架角度为180°的射野(记为G180),生成验证计划,在Varian Clinac i X加速器上实施,使用二维电离室矩阵Matri XX进行测量。通过Omni Pro I’m RT软件,模拟绝对剂量的偏移(记为d),d在-10%~10%取值,间隔1%,观察相应Matri XX剂量验证结果变化。结果总体上看,绝对剂量向正负方向偏移时,Gamma通过率下降,平均Gamma值变大。本研究中,Gamma通过率的最大值并不在d=0处,而在d处于-3%^-1%之间;平均Gamma值最小点在d处于-3%~0之间。结论绝对剂量的偏移对调强计划射野的Matri XX验证结果有一定影响,在分析Matri XX剂量验证结果时,应将绝对剂量的偏移因素纳入考虑中,同时在放疗日常工作中,应定期检查校准加速器的绝对剂量。 Objective To discuss the influence of deviation in absolute dose on verification results of IMRT fields of MatriXX.Methods Fields with gantry angle180°(G180)were selected for24patients with cervical cancer,and corresponding verificationplans were created.The verification plans were implemented in the Varian Clinac iX linear accelerator,and the dose distributionswere measured by the two-dimensional ionization chamber array(MatriXX).OmniPro I’mRT software was used to simulatethe deviation in absolute dose(d),and d ranged from-10%to10%with interval of1%.Corresponding dose verification resultswith different values of d were analyzed.Results By and large,gamma passing rate was decreased and value of average gammawas increased when absolute dose deviated to positive or negative directions.The maximum gamma passing rate was not in theposition with d=0,but in the range of-3%to-1%,and the values of d were in the range of-3%to0when average gamma achievedthe minimum value.Conclusion Deviation in absolute dose has a significant influence on verification results of IMRT fields ofMatriXX.When analysing verification results,the factor of deviation in absolute dose should be taken into consideration.In routinework of radiotherapy,the absolute dose of Clinac accelerator should be calibrated regularly.
作者 柏正璐 时飞跃 秦伟 蒋红兵 BAI Zhenglu;SHI Feiyue;QIN Wei;JIANG Hongbing(Department of Medical Equipment,Nanjing Medical University Affiliated Nanjing Hospital (Nanjing First Hospital), Nanjing Jiangsu 210006, China;Department of Radiation Oncology, Nanjing Medical University Affiliated Nanjing Hospital (Nanjing First Hospital), Nanjing Jiangsu 210006, China;Nanjing Health Information Center, Nanjing Jiangsu 210003, China)
出处 《中国医疗设备》 2017年第10期48-51,共4页 China Medical Devices
基金 南京市医学科技发展资金"青年工程"人才培养专项经费资助(QRX11033)
关键词 Gamma通过率 绝对剂量 剂量验证 宫颈癌 调强放疗 Gamma passing rate absolute dose dose verification cervical cancer intensity modulated radiation therapy
  • 相关文献

参考文献9

二级参考文献72

共引文献34

同被引文献62

引证文献6

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部