摘要 目的探讨POCT血糖仪与全自动生化分析仪检测结果的相关性,定期评价便携式血糖仪的可靠性.方法纵向收集曲靖市第一人民医院内分泌科住院的糖尿病或非糖尿病患者,2012年1月至2015年12月定期每年2次的血糖比对值,POCT血糖值与罗氏7600全自动分析仪的检测结果,共448例,采用SPSS进行统计学处理.结果内分泌代谢科POCT血糖仪与罗氏7600全自动分析仪的检测结果,相关系数r>0.95,相关性好,偏倚符合卫生部要求.结论 POCT血糖仪简便、快速、准确、临床实用性好,内分泌代谢科POCT血糖仪的检测准确性、重复性、稳定性好. Objective To investigate the correlation between POCT glucose meter and automatic biochemical analyzer test results,to evaluated the reliability of portable POCT blood glucose meter.Methods We vertically collected the blood glucose values that diabetics and non diabetics hospitalized in The First People's Hospital of Qujing of Endocrinology,from January2012to December2015regularly twice a year,the POCT blood glucose meter and Roche7600automatic analyzer test results,all of448cases,using SPSS17.0for statistical analysis.Results The correlation of POCT blood glucose meter test results with Roche7600automatic analyzer was coefficiently,r>0.95,meeting the standards of the Ministry of Public Health.Conclusions POCT glucose meter is simple,rapid,accurate and clinically useful.POCT glucose meter in our department,has good detection accuracy,stability and repeatability.
作者 李贵梅 陈卫文 徐应波 殷晓娥 LI Gui-mei;CHEN Wei-wen;XU Ying-bo;YIN Xiao-e(The First People’s Hospital of Qujing,Qujing Yunnan 655000,China)
出处 《昆明医科大学学报》 CAS 2017年第8期106-109,共4页 Journal of Kunming Medical University
基金 云南省科技厅-昆明医科大学应用基础研究联合专项基金资助项目(2014FZ060)
关键词 POCT血糖仪检测 自动生化分析仪检测 纵向 相关性 POCTglucometer detection Automatic biochemical analyzer detection Vertical Correlation
  • 相关文献

参考文献21

二级参考文献160

共引文献178

同被引文献131

引证文献20

二级引证文献53

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部