摘要 目的评价新一代乳胶增强免疫比浊法抗环瓜氨酸肽抗体(Anti-CCP)试剂盒在日立7600-020全自动生化分析仪上的试剂性能。方法根据美国临床和实验室标准化协会(CLSI)相关文件,评价Anti-CCP试剂盒的准确度、批内精密度、批间精密度、线性范围,并分别与雅培、罗氏的检测方法作比较,考察其与目前较主流检测方法间的阴阳性符合率。结果 Anti-CCP试剂盒检测准确度符合要求。质控水平1批内和批内精密度分别为2.06%和2.43%,质控水平2批内和批间精密度分别为1.71%和1.83%。线性范围(0~980 U/mL)符合要求。与雅培化学发光和罗氏电化学发光检测系统阳性符合率分别为86.7%和78.1%、阴性符合率分别为92.1%和94.9%,均大于60%。结论 Anti-CCP试剂盒在日立7600-020全自动生化分析仪上的试剂性能能够满足临床应用的要求,适合于检验科常规测定。
作者 周立红
出处 《国际检验医学杂志》 CAS 2017年第24期3489-3490,共2页 International Journal of Laboratory Medicine
  • 相关文献

参考文献3

二级参考文献17

  • 1Mediwake R, Isenberg DA, Schellekens GA, et al. Use of anti citrullinatce peptide and anti RA33 antibodies in distinguishing erosive arthritis in patients with systemic lupus erythematosus and rheumatoid arthritis[J]. Ann Rheum Dis, 2001,60(i) :67-68.
  • 2Schellekens GA, Visser H, De Jong BAW, et al. The diagnostic properties of rtheumatoid arthritis antibodies recognizing a cyclic cirtrullinated peptide[J].Arthritis Rheum, 2000,43(12): 155-163.
  • 3Vincent C, Simon M, Sebbag M, et al. Immunoblotting detection of autoantibodies to human epidermis filaggein: a new diagnostic test for rheumatoid arthritis[J].J Rheumatol, 1998,25(5) : 838-846.
  • 4Francesco Dati,Jillian R. Tate,Santica M. Marcovina,Armin Steinmetz.First WHO/IFCC International Reference Reagent for Lipoprotein(a) for Immunoassay – Lp(a) SRM 2B[J].Clinical Chemistry and Laboratory Medicine.2004(6)
  • 5J. C. Hopewell,U. Seedorf,M. Farrall,S. Parish,T. Kyriakou,A. Goel,A. Hamsten,R. Collins,H. Watkins,R. Clarke.Impact of lipoprotein(a) levels and apolipoprotein(a) isoform size on risk of coronary heart disease[J]. J Intern Med . 2014 (3)
  • 6Loscalo J,Fless GM,et al.Lipoprotein (a): A uniquerisk factor for atherothrombotic disease. Arteriosclerosis . 1990
  • 7The National Conmmittee for Clinical Laboratory Standards.Method comparison and bias estimation using patient samples. NCCLS Document EP9-A2 . 2002
  • 8How to define and determine reference intervals in the clinical laboratory. NCCLS C28-A2 . 2000
  • 9Glick M R,Ryder K W,Jackson S A.Graphical comparisons of interferences in clinical chemistry instrumentation. Clinical Chemistry . 1986
  • 10Clinical and Laboratory Standard Institute.User verification of performance for precision and trueness. EP15-A2 . 2005

共引文献11