不同管电流技术对胸部CT健康体检图像质量的影响及辐射剂量的优化 被引量:2

Study on image quality of chest CT physical examination with different tube current techniques,and radiation dose optimization
下载PDF
导出
摘要 目的探讨胸部CT健康体检达到诊断要求且辐射剂量最低的管电流。方法将50例健康体检者随机分5组,每组10例:试验组设立管电流梯度分别为20 mA、30 mA、50 mA,自动mA的四个组;对照组为220mA。分别对上述5组的纵隔窗及肺窗进行质量评定。对试验组及对照组进行t检验,试验组间采用方差分析。找出符合诊断要求及辐射剂量最低的管电流。结果试验组4个管电流梯度的容积剂量指数(Ctdivol)剂量长度值(Dlp),有效吸收剂量(ED)均低于对照组。经统计学分析,二者差异均有统计学意义(P<0.05)。试验组中以20 mA组的Ctdivol、Dlp、ED最低。试验组4个剂量梯度(20 mA、30 mA、50 mA、自动mA组)的纵隔窗及肺窗的质量评分分别为:1.99±0.31/2.88±0.62,2.49±0.52/3.21±0.54,3.55±0.65/4.04±0.63,4.97±0.12/4.97±0.12。当纵隔窗及肺窗大于3时定义为可以达到诊断要求,因此管电流30 mA时被认为是可以作为胸部低剂量CT体检的最佳管电流。结论管电流30 mA时可以作为胸部低剂量CT体检的最低剂量值。 Objective To investigate the lowest dose of the tube current on the chest CT physical examination which can satisfy the diagnostic requirements.Methods The 50 persons subjected to physical examination were randomly divided into 5 groups,with an average of 10 persons:the 4 trial groups were set up 4 dose gradient,they were 20mA,30mA and 50mA,and automatic mA respectively.Control group was set as 220mA.The quality of the mediastinum window and lung window in above 5 groups were evaluated respectively.The t test was performed in trial groups and control group,and the analysis of variance was used among trial groups.Then the lowest radiation dose were seeked which dose can meets the diagnostic requirements.Results The volume dose index(Ctdivol),dose length product(Dlp)and effective absorbed dose(ED)of the 4 dose gradients in trial groups were lower than those in control group.The both's differences were statistically significant(P<0.05).The Ctdivol,Dlp and ED in the 20mA group were the lowest in trial group.Trial group received 4 doses(20mA,30mA,50mA gradient,automatic mA group)quality score of the mediastinal window and pulmonary window respectively:1.99±0.31/2.88±0.62,2.49±0.52/3.21±0.54,3.55±0.65/4.04±0.63,4.97±0.12/4.97±0.12.When the mediastinum window and the lung window were larger than 3,defined as the diagnostic requirements,the 30mA group was considered to be the best tube current for chest low dose CT examination.Conclusion When the tube current is 30MA,it can be used as the lowest dose for chest low dose CT examination.
作者 吴振东 李杰 孙宗琼 岳建国 WU Zhen-dong;LI Jie;SUN Zong-qiong(Image Dept.,the 4th People's Hospital of Wuxi City,Wuxi,Jiangsu 214062,China)
出处 《实用医药杂志》 2018年第2期119-121,共3页 Practical Journal of Medicine & Pharmacy
基金 江苏省卫生和计划委员会课题(Q201615)
关键词 低剂量 管电流 胸部 螺旋CT Low dose Tube current Chest Spiral CT
  • 相关文献

参考文献4

二级参考文献35

  • 1李剑颖.GE VCT——高图像质量和低剂量的和谐[J].CT理论与应用研究(中英文),2006,15(3):57-60. 被引量:5
  • 2刘建新,邓刚,尹晓明.低剂量螺旋CT在胸部体检中应用价值探讨[J].医用放射技术杂志,2006(8):52-54. 被引量:3
  • 3Itoh S,Ikeda M,Schweden F,et al.Lung cancer screening:minimum tube current required for helical CT.Radiology,2000,215(1):175-183.
  • 4Brener DJ,Elliston CD,Hall EJ,et al.Estimated risks of radiation-induced fatal cancer from pediatric CT.AJR Am J Roentgenol,2001,176(2):289-296.
  • 5Hojreh A,Kainberger F,Puig S.Low dose multislice CT in the pediatric patient.Radiology,2003,43(12):1051-1055.
  • 6Mettler FA Jr,Wiest PW,Locken JA,et al.CT scanning:patterns of use and dose.J Radiol Prot,2000,20(4):353-359.
  • 7Greess H,Nmayr A,Wolf H,et al.Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose).Eur Radiol,2002,12(6):1571-1576.
  • 8Prasad SR,Wittram C,Shepard JA,et al.Standard-dose and 50%-reduced-dose chest CT:comparing the effect on image quality.AJR Am J Roentgenol,2002,179(2):461-465.
  • 9中島正俊.撮影条件自動適正化機能を用いたル-チン化への検討.日本放射線技術学会雑誌,2002,58(4):502-510.
  • 10Kalra MK,Rizzo S,Maher MM,et al.Chest CT performed with z-axis modulation:scanning protocol and radiation dose.Radiology,2005,237(1):303-308.

共引文献94

同被引文献16

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部