第三代双源CT不同扫描模式抑制下颈部伪影的对比

Comparison on the suppression of the artifacts in lower neck artery examination by the third-generation dual-source CT in different scanning mode
下载PDF
导出
摘要 目的:通过对比下颈部CT常规扫描、双能量模式扫描、无预设值CARE kV联合CARE Dose 4D自动控制曝光技术,探索下颈部CT扫描抑制伪影的个性化方案.方法:随机选取进行下颈部CT检查的患者99例,分成3组,A组:固定千伏、毫安扫描;B组:无预设值CARE kV联合CARE Dose 4D自动控制曝光技术扫描;C组:双能量模式扫描,分别将图像质量评分、噪声值、信噪比及容积CT剂量指数进行对比.结果:C组图像质量评分(3.70±0.51)高于B组(3.36±0.50),P<0.05,B组图像质量评分(3.36±0.50)明显高于A组(3.00±0.59),P<0.01;B组容积CT剂量指数(6.38±1.40)mGY明显低于A组(9.29 mGY),P<0.01;A组容积CT剂量指数(9.29mGY)明显低于C组(12.58±4.74)mGY,P<0.01;C组噪声值(7.51±2.29)明显低于A组(12.03±3.52)、B组(12.18±4.47),两者P<0.01;C组信噪比(7.57±2.20)明显高于A组(5.19±1.86)、B组(5.51±1.86),两者P<0.01;A、B组噪声值、信噪比无明显统计学意义(分别为P=0.86,P=0.52).结论:建议常规下颈部CT扫描使用无预设值CARE kV联合CARE Dose 4D自动控制曝光技术,可获得辐射剂量最低、图像质量较优的影像,而双能量扫描模式能更大程度抑制下颈部伪影,在诊断下颈部疑难病变、定位细微病灶时,或在征得病人同意的情况下,可使用双能量模式获得更高质量的影像. Objective:To explore a an individualized program for suppressing artifact in CT scan of the cervix through comparing the difference of the CT routine scan,dual energy mode scan,the non-presetting intelligent optimal tube voltage scanning(CARE KV)combined with automatic tube current modulation(CARE Dose 4D)technology on the cervix.Methods:Nighty-nine patients received the cervix CT examination in our hospital were randomly selected and divided into 3 groups,33 cases each.Patients in group A received the conventional fixed kilovolt and milliampere CT scan,and patients in group B received the non-presetting intelligent optimal tube voltage scanning(CARE KV)combined with automatic tube current modulation(CARE Dose 4D)technology,and patients in group C received dual energy mode CT scan.Eventually the image quality(noise,SNR,subjective ratings)and radiation dose(CTDI vol)of the three groups were compared.Results:The image quality score of group C(3.70±0.51)was higher than that of group B(3.36±0.50)(P<0.05),and the image quality score of group B(3.36±0.50)was higher than that of group A(3.00±0.59)(P<0.01).The CT Volume dose index of group B(6.38±1.40 mGY)was lower than that of group A(9.29 mGY)(P<0.01),and the CT volume dose index of group A(9.29 mGY)was lower than that of group C(12.58±4.74 mGY)(P<0.01).The noise of group C(7.51±2.29)was obviously lower than that of group A(12.03±3.52)and group B(12.18±4.47)(P<0.01),while the SNR between group C(7.57±2.20)was obviously higher than that of group A(5.19±1.86)and group B(5.51±1.86)(P<0.01).The noise and SNR of group A and B showed no significant difference(group A P=0.86,group B P=0.52).Conclusion:It suggests that routine cervix scan using the non-presetting intelligent optimal tube voltage scanning(CARE KV)combined with automatic tube current modulation(CARE Dose 4D)technology in order to obtain the image with the lowest radiation dose and the higher quality.While diagnosing perplexing neck lesions and locating small lesions,dual energy mode scan can be used to obtain image of even higher quality with the consent of the patients.
作者 蒋博 梅习龙 孙剑宁 黄艳芳 吴雄 樊丽丹 周湘乐 邓凯 JIANG Bo;MEI Xilong;SUN Jianning;HUANG Yanfang;WU Xiong;FAN Lidan;ZHOU Xiangle;DENG Kai(Department of Radiology,the Second Xiangya Hospital,Central South University,Changsha 410011,China)
出处 《暨南大学学报(自然科学与医学版)》 CAS CSCD 北大核心 2018年第2期137-142,共6页 Journal of Jinan University(Natural Science & Medicine Edition)
基金 国家自然科学基金项目(81671671)
关键词 颈部 双能量CT 肩关节 伪影 辐射剂量 噪声 cervical dual energy CT shoulder artifact radiation dose noise
  • 相关文献

参考文献9

二级参考文献66

  • 1吴晓华,马大庆,张忠嘉,冀景玲,张岩松.多层螺旋CT胸部低剂量扫描发现肺结节的临床研究[J].中华放射学杂志,2004,38(7):767-770. 被引量:86
  • 2王晓峰,彭明辰.关于多排螺旋CT剂量的探讨[J].医疗设备信息,2004,19(12):23-25. 被引量:20
  • 3殷瑞根,吴树春,王冬青,彭卫斌,孙维斌,范志刚,廖谦,赵亮.低剂量多层螺旋CT在急腹症中的初步应用[J].临床放射学杂志,2005,24(7):609-612. 被引量:29
  • 4周诚,沈云.多层螺旋CT的图像重组原理和技术进展[J].中华放射学杂志,2006,40(9):991-993. 被引量:49
  • 5Barrett J F,Keat N.Artifacts in CT:recognition and avoidance[J].Radiographics,2004,24(.):1679-1691.
  • 6Brink M,de Lange F,Oostveen L J,et al.Arm raising at exposure controlled multi-detector trauma Ctof thoracoabdominal region:higher imagequality,lower radiation dose[J].Radiology,2008,249:661-670.
  • 7Greess H,Lutze J,Nomary A,et al.Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation[J].Eur Radiol,2004,14:995-999.
  • 8Slovis TL.CT and computed radiography:the pictures aregreat,but is the radiation dose greater than required?[J].Am JRoentgenol,2002,179(1):39-41.
  • 9Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radiationdose optimization[J].Radiology.2004,230(3):619-628.
  • 10Brix G,Lechel U,Petersheim M,et al.Dynamic contrast-enhancedCT studies:balancing patient exposure and image noise[J].InvestRadiol.2011,46(1):64-70.

共引文献106

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部