期刊文献+

生物可吸收镁合金支架治疗20例兔模型颈内动脉瘤影像学和组织病理学研究 被引量:1

Bioabsorbable magnesium alloy stent in the treatment of internal carotid aneurysms:an angiographic and histopathology study of 20 rabbit model
下载PDF
导出
摘要 目的构建生物可降解镁合金覆膜支架植入兔颈内动脉瘤模型,评价生物可降解镁合金覆膜支架在兔颈内动脉瘤模型的组织学反应,支架降解和内皮化情况。方法选取20只新西兰白兔构建颈内动脉瘤模型,通过外科手术于72 h内植入可降解镁合金覆膜支架。不同时间段行DSA血管造影检查评价载瘤动脉是否通畅及再狭窄率等情况。随访4个月,光镜和扫描电镜观察包括6和12个月的颈内动脉造影和血管内皮化及临床评估。结果共成功构建20例新西兰大白兔颈动脉瘤模型,顺利植入生物可降解镁合金覆膜支架。术后立即行血管造影,显示18例完全闭塞,仅2例动脉瘤腔内有部分血栓形成。光镜扫描下所有动脉瘤腔内均可见血栓形成,有些甚至发生机化。1周血管造影显示直段血管侧壁动脉瘤保持很好通畅,支架释放12个月后动脉瘤腔完全隔离。钼靶检测结果显示,镁合金覆膜支架在1 a后明显降解,期间未发现支架移位。术后1个月血管出现内皮化进程,术后6个月完成内皮化过程。结论应用可降解镁合金覆膜支架有助于治疗颈内动脉瘤。支架植入后颈动脉的通畅率得到了显著提高。镁合金覆膜支架具有可生物降解特性,随着时间的延长而降解,于术后6个月完成内皮化。 Objective To establish a biodegradable magnesium alloy stent graft for internal carotid artery aneurysms in rabbits,and to evaluate the histological response,stent degradation and endothelialization of the biodegradable alloy covered stent in the rabbit internal carotid artery aneurysm model.Methods Twenty New Zealand white rabbits were selected to establish rabbit models of internal carotid artery aneurysm by surgical operation.After implantation,a biodegradable magnesium alloy stent was implanted within 72 hours.DSA angiography was performed at different times to evaluate whether the carrying artery was smooth and restenosis rate.The endothelialization of the aneurysm cavity and blood vessel was observed by light and scanning electron microscopy.Results Biodegradable magnesium alloy covered stents were successfully implanted in 20 New Zealand white rabbits with the lateral common carotid artery aneurysm model.The model of internal carotid aneurysm maintained a good open cavity.Only 2 of the aneurysms were formed in the lumen of the aneurysm,and the common carotid artery was unobstructed.In all aneurysms,thrombosis was seen in all aneurysms,and some were even computerized.One week after implantation,angiography showed that the lateral wall aneurysm of the direct segment remained smooth,and the aneurysm cavity was completely isolated after the stent was released for 12 months.One year after implantation,molybdenum target detection showed the coated stent of magnesium alloy was degraded obviously and the contour disappeared.The endothelialization process began at 1 month after operation,and the endothelialization process was completed 6 months after operation.Conclusion The application of degradable magnesium alloy coated stent in the treatment of internal carotid artery aneurysm is very feasible.The patency rate of carotid artery stenting has been significantly improved.The magnesium alloy coated stent has the characteristics of biodegradation,which degraded with time and completed endothelialization at 6 months after the operation.
作者 张宪亮 崔红凯 Zhang Xianliang;Cui Hongkai(Department of Intervention,Central Hospital of Zhoukou City,Zhoukou 466000,China;Department of Intervention,the First Affiliated Hospital of Xinxiang Medical University,Weihui 453100,China)
出处 《河南医学研究》 CAS 2018年第5期781-785,共5页 Henan Medical Research
基金 河南省医学科技攻关计划项目(201301006)
关键词 颈内动脉瘤 动物模型 镁合金覆膜支架 内皮化 rabbit internal carotid aneurysm animal model magnesium alloy covered stent endothelialization
  • 相关文献

参考文献6

二级参考文献96

  • 1高家诚,李龙川,王勇,乔丽英.表面改性纯镁的细胞毒性和溶血率[J].稀有金属材料与工程,2005,34(6):903-906. 被引量:26
  • 2谭小伟,高家诚,王勇,乔丽英,伍沙.医用纯镁的热处理试验研究[J].科技导报,2006,24(2):67-69. 被引量:9
  • 3耿芳,谭丽丽,张炳春,郑丰,杨柯.多孔镁作为新型骨组织工程材料的研究探索[J].材料导报,2007,21(5):76-78. 被引量:24
  • 4高家诚,伍沙,乔丽英,王勇.镁及镁合金在仿生体液中的腐蚀降解行为[J].中国组织工程研究与临床康复,2007,11(18):3584-3586. 被引量:25
  • 5Waksman R,Pakala R,Kuchulakanti PK,et al.Safety and efficacy of bioabsorbale magnesium alloy stents in porcine coronary arteries[J].Catheter Cardiovas Interv,2006,68:607-617.
  • 6Virmani R,Farb A,Guagliumi G,et al.Drug-eluting stents:caution and concerns for long-term outcome[J].Coron Artery Dis,2004,15:313-318.
  • 7Hehrlein C.Promises of biodegradable stents[J].Catheter Cardiovasc Interv,2007,69:739.
  • 8Hansi C,Arab A,Rzany A,et al.Differences of platelet adhesion and thrombus activation on amorphous silicon carbide,magnesium alloy,stainless steel,and cobalt chromium stent surfaces[J].Catheter Cardiovasc Interv,2009,73:488-496.
  • 9Waksman R,Erbel R,Di Mario C,et al.Early-and long-term intravascular ultrasound and angiographic findings after bioabsorbable magnesium stent implantation in human coronary arteries[J].JACC Cardiovasc Interv,2009,2:312-320.
  • 10Ghimire G,Spiro J,Kharbanda R,et al.Initial evidence for the return of coronary vasoreactivity following the absorption of bioabsorbable magnesium alloy coronary stents[J].Eurointervention,2009,4:481-484.