扫描参数变化对FLAIR序列金属伪影的影响 被引量:3

Influence of Change of Scanning Parameters on Metal Artifacts in FLAIR Sequence
下载PDF
导出
摘要 目的探讨不同扫描参数对FLAIR序列金属伪影的影响,以通过调整扫描参数来减小金属伪影。方法选用镍铬合金制成左下6标准桩核冠模型。应用1.5 T MRI扫描仪对桩核冠模型行横断面FLAIR扫描,扫描参数为西门子1.5T MR扫描仪的推荐值。保持其它参数与标准扫描序列参数一致,逐一改变以下各参数进行成像:TR、TE、FOV、Matrix、BW、NEX、ETL、Thk、FA。确定伪影最大的层面并测量伪影面积。应用SAS V8统计学软件进行统计学分析。结果调整不同FOV值,不同Matrix值及FA值,镍铬合金伪影大小差异有统计学意义(F值分别为F=67.04,F=48.75,F=5.61,P<0.05)。结论在FLAIR序列中,可以通过减小视野、增加矩阵,增大翻转角减小金属伪影。 Objective To discuss the influence of different scanning parameters on the metal artifacts in FLAIR sequence,in order to reduce the artifacts by adjusting the scanning parameters.Methods Ni-Cr alloy were molded into standard post-core-crown posterior tooth model.The model were scanned by 1.5T MR scanner.Axial FLAIR sequence was scanned.The scan parameters was the recommended values for the SIEMENS 1.5T MR scanner.Changing the following parameters:TR,TE,FOV,Matrix,BW,NEX,ETL,Thk,FA one by one without affecting the image quality.To determine the section of the largest artifacts and measure the area.Statistical testing was performed.Results There was statistically significant differences of metal artifacts in different FOV values(F=67.04,P<0.001).There was statistically significant differences of metal artifacts in different Matrix values(F=48.75,P<0.001).There was statistically significant differences of metal artifacts in different FA values(F=5.61,P=0.04).Conclusion In FLAIR sequence,decreasing the FOV,increasing Matrix and FA can decrease the artifacts.
作者 李晖 陈一 王雅静 陈伟彬 王志强 代振杰 LI Hui;CHEN Yi;WANG Ya-jing(Department of Radiology,North China University of Science and Technology Affiliated Hospital,Tangshan 063000,Hebei Province,China)
出处 《中国CT和MRI杂志》 2018年第3期137-139,共3页 Chinese Journal of CT and MRI
基金 河北省科技支撑计划(1627 77117D) 河北省卫生厅青年科技课题科研基金项目资助(20150856)
关键词 镍铬合金 伪影 磁共振成像 液体衰减反转恢复 Ni-Cr Alloy Artifacts Magnetic Resonance Imaging(MRI) Fluid Attenuated Inversion Recovery(FLAIR)
  • 相关文献

参考文献9

二级参考文献69

  • 1刘志强,马得廷,郭岩,孔庆奎.FLAIR序列在脑室系统中的应用[J].中国CT和MRI杂志,2005,3(1):11-13. 被引量:8
  • 2陈渊博,陈春晓,周卉芬.磁共振成像中伪影的研究[J].国际生物医学工程杂志,2006,29(4):214-218. 被引量:12
  • 3Hajnal J V, Bryant D J, Kasuboski L,et al. Use of fluid attenuated inversion recovery(FLAIR) pulse sequences in MR1 of the brain [ J ]. J Comput Assist Tomogr, 1992,16 (6) :841 - 844.
  • 4Hasherm R H, Brown J J, Hesselink J R, et al. MR and CT of lacunar infrarcts [ J ]. Radiology, 1995,196 (2) :505 - 510.
  • 5Nakada T. Clinical experience on 3.0 T systems in Niigata, 1996 to 2002. Invest Radiol, 2003, 38(7): 377-384.
  • 6Nael K, Fenchel MC, Kramer U, et al. Whole-body contrast-enhanced magnetic resonance angiography: new advances at 3.0 T. Top Magn Reson Imaging, 2007, 18(2): 127-134.
  • 7Nakada T. Clinical application of high and ultra high-field MRI. Brain & Development, 2007, 29(6): 325-335.
  • 8Yablonskiy DA, Ha-acke EM. Theory of NMR signal behavior in magnetically inhomogeneous tissues: the static dephasing regime. Magn Reson Med, 1994, 32(6): 749-763.
  • 9Bemstein MA, Huston III J, Ward HA. Imaging artifacts at 3.0 T. J Magn Reson Imaging, 2006, 24(4): 735-746.
  • 10Merkle EM, Dale BM, Thomas J, et al. MR liver imaging and cholangiography in the presence of surgical metallic clips at 1.5 and 3 T. Eur Radiol, 2006, 16(10): 2309-2316.

共引文献60

同被引文献43

引证文献3

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部