摘要
为了更好地测量液体折射率,提出了一种基于表面等离子共振波长测量液体折射率的方法,采用Kretschman结构建立了模型,进行了软件仿真,并搭建实验平台进行了实验研究,分析了实验和理论之间的误差来源。结果表明,当折射率在1.33RIU^1.36RIU的范围内变化时,表面等离子共振吸收峰随液体样品折射率的变化产生了频移,其灵敏度可达4808.94nm/RIU。该方法可以准确测量液体的折射率,且系统结构简单,具有较高的灵敏度。
In order to measure refractive index of liquid more accurately,one method based on surface plasma resonance wavelength was proposed.The model was built by using Kretschman structure and software simulation was carried out.The experimental platform was set up and the experimental research was carried out.The error between experimental and theoretical results was analyzed.The results show that,when the refractive index changes within the range of 1.33RIU^1.36RIU,the absorption peak of surface plasma resonance produces a frequency shift with the change of refractive index of liquid sample.And its sensitivity is up to 4808.94nm/RIU.Refractive index of liquid can be measured accurately with the method and the system has simple structure and high sensitivity.
作者
邹璐
李慧明
菅傲群
王雷阳
ZOU Lu;LI Huiming;JIAN Aoqun;WANG Leiyang(MicroNano System Research Center,Taiyuan University of Technology,Taiyuan 030024,China;Key Laboratory of Advanced Transducers and Intelligent Control System,Shanxi Province and Ministry of Education,Taiyuan University of Technology,Taiyuan 030024,China)
出处
《激光技术》
CAS
CSCD
北大核心
2018年第3期357-361,共5页
Laser Technology
基金
国家自然科学基金资助项目(61501316
61471255
61474079
51505324
51622507)
山西省基础研究青年基金资助项目(2015021092)
山西省归国留学基金资助项目(2015-047)
国家高技术研究发展计划资助项目(2015AA042601)
山西省优秀人才技术创新基金资助项目(201605D211027)
关键词
传感器技术
共振波长
表面等离子体共振
灵敏度
sensor technique
resonant wavelength
surface plasma resonance
sensitivity