两套检测系统糖化血红蛋白检测结果的一致性分析 被引量:3

Consistency Analysis of the Results of Glycosylated Hemoglobin Detection in Two Detection Systems
下载PDF
导出
摘要 目的探讨两套检测系统糖化血红蛋白(HbA1c)检测结果的一致性。方法按照EP9-A2文件的要求,以PRIMUS Ultra2 HbA1c分析仪为参比系统,PRIMUS PDQ-Plus HbA1c分析仪为待评系统,使用新鲜抗凝全血对两套检测系统的结果进行比对分析,并评估其在不同医学决定水平(XC)处的相对偏倚(SE%)。结果两套检测系统HbA1c检测结果的回归方程为Y=0.9927X+0.1208,相关系数r^2=0.9944,不同XC处的SE%均小于1/2TEa,能被临床所接受。结论 PRIMUS PDQ-Plus与PRIMUS Ultra2 HbA1c分析仪检测结果具有良好的一致性,可使用相同的参考范围,能满足临床对糖尿病患者的诊断、治疗和监测需求。 Objective To investigate the two detection system of glycosylated hemoglobin(HbA1c)the consistency of test results.Methods According to the requirements of the EP9-A2 documents,the PRIMUS Ultra2 HbA1c analyzer as the reference system,the PRIMUS PDQ-Plus HbA1c analyzer for evaluation system,the use of fresh anticoagulant blood were compared and analyzed the results of two detection system,and evaluate the medical decision level in different relative bias(XC)at(SE%).Results Two the regression equation of HbA1c detection system was Y=0.9927X+0.1208,correlation coefficient 2=0.9944,and SE%at different XC were less than 1/2TEa,which could be accepted by clinical practice.Conclusion The results of PRIMUS PDQ-Plus and PRIMUS Ultra2 HbA1c analyzer have good consistency and can be used in the same reference range.It can meet the needs of clinical diagnosis,treatment and monitoring of diabetic patients.
作者 罗映杰 林凤 刘艳婷 肖光军 LUO Ying-jie;LIN Feng;LIU Yan-ting;XIAO Guang-jun(Department of Laboratory,Suining Central Hospital,Suining 629000,Sichuan,China;Major in Medical Laboratory Technology,Ya'an Vocational and Technical College,Ya'an 625000,Sichuan,China)
出处 《医学信息》 2018年第1期74-76,共3页 Journal of Medical Information
关键词 糖化血红蛋白分析仪 比对分析 偏倚评估 一致性 Glycosylated hemoglobin analyzer Comparison analysis Bias assessment Consistency
  • 相关文献

参考文献6

二级参考文献95

  • 1顾燕云,邢慧莉,盛正妍,伍贻新,罗敏.酶法测定糖化血清白蛋白的多中心临床研究[J].上海医学,2006,29(7):432-434. 被引量:31
  • 2中国合格评定国家认可委员会.已获认可机构名称[EB/OL].[2010-12-29]http://219.238.178.49/Acc_Search2.asp?Class=L.
  • 3International Expert Committee. International Expert Commit- tee report on the role of the AIC assay in the diagnosis of dia- betes. Diabetes Care, 2009, 32: 1327-1334.
  • 4World Health organization. Use of glycated haemoglobin(HbA1 c)in the diagnosis of diabetes mellitus: abbreviated report of a WHO consultation. Geneva: WHO, 2011.
  • 5IFCC. Recommendation for term and measurement unit for "HbA1 c'. Clin Chem Lab Med, 2007, 45 : 1081-1082.
  • 6Miedema K. Towards worldwide standardisation of HbA1 c de- termination. Diabetologia, 2004, 47:1143-1148.
  • 7Hanas R, John G. International HbA1 c Consensus Commit- tee. 2010 Consensus statement on the worldwide standardiza- tion of the hemoglobin AIC measurement. Clin Chem Lab Med, 2010, 48:775-776.
  • 8Sacks DB, Arnold M, Bakris GL, et al. Position statement ex- ecutive summary: Guidelines and recommendations for labora- tory analysis in the diagnosis and management of diabetes mel- litus. Diabetes Care, 2011, 34:1419-1423.
  • 9Goodall I, Colman PG, Schneider HG, et al. Desirable per- formance standards for HbA1 c analysis precision, accuracy and standardization. Clin Chem Lab Med, 2007, 45:1083-1097.
  • 10Molinaro RJ. Targeting HbA1 c: standardization and clinical la- boratory measurement. Mlo Med Lab Obs, 2008, 40:10-19.

共引文献6925

同被引文献14

引证文献3

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部