期刊文献+

彩色角膜接触镜染色位置判断方法的研究

Research on Evaluating Method of Location of Dyeing Materials for Color Contact Lenses
下载PDF
导出
摘要 该文研究了结合使用冰冻切片及扫描电镜试验对彩色角膜接触镜进行纵断面及表面形貌观察,对产品的染色位置进行评价,形成了一种判断彩色角膜接触镜染色位置有效的试验方法,并使用该方法对60批产品染色位置进行评价。结果表明,60批产品仅有7批染色颗粒完全位于产品内部,该方法可以实现彩色角膜接触镜染色位置的有效判断。 Combining frozen section and scanning electron microscopy experiments to observe the cross section and surface state of the samples,this study finds an effective method to evaluate the location of dyeing materials of color contact lenses.Sixty samples were evaluated on their dyeing location statistically using this method.The results showed that there were only 7 lots of samples with their dyeing materials within their product.This method is effective in evaluating the location of dyeing materials for color contact lenses.
作者 张庆 万敏 王常斌 ZHANG Qing;WAN Min;WANG Changbin(Physical Department,Shandong Quality Inspection Center for Medical Devices,Jinan,250101)
出处 《中国医疗器械杂志》 2018年第4期296-298,共3页 Chinese Journal of Medical Instrumentation
基金 山东省食品药品监督管理局基金资助项目(SF1406014002)
关键词 彩色角膜接触镜 染色位置 角膜感染 冰冻切片 扫描电镜 color contact lenses dyeing location corneal infection frozen section scanning electron microscopy
  • 相关文献

参考文献1

二级参考文献11

  • 1齐备.隐形眼镜手册[M]上海:上海科学技术出版社,1998252-254.
  • 2杨熹明.台湾市售美妆隐形眼镜之色素结构调查[R],2010.
  • 3Zhao Z,Carnt N A,Aliwarga Y. Care regimen and lens material influence on silicone hydrogel contact lens deposition[J].Optometry and Vision Science,2009,(03):251-259.doi:10.1097/OPX.0b013e318196a74b.
  • 4Sauer A,Bourcier T. Microbial keratitis as a foreseeable complication of cosmetic contact lenses:a prospective study[J].Acta Ophthalmologica,2011,(05):439-442.doi:10.1111/j.1755-3768.2011.02120.x.
  • 5Van Zyl LM,Cook C. Bacterial keratitis and corneal scarring secondary to cosmetic contact lens wear[J].South African Medical Journal,2010,(01):37-38.
  • 6Rakow P L. J Contact lenses meet cosmetic needs of patients[J].Ophthalmic Nurs Technol,1999,(04):177-179.
  • 7Hong S,Lee J R,Lim T. Pigment deposition of cosmetic contact lenses on the cornea after intense pulsed-light treatment[J].Korean Journal of Ophthalmology,2010,(06):367-370.doi:10.3341/kjo.2010.24.6.367.
  • 8Zhao Z,Naduvilath T,Flanagan J L. Contact lens deposits,adverse responses,and clinical ocular surface parameters[J].Optometry and Vision Science,2010,(09):669-674.doi:10.1097/OPX.0b013e3181ea1848.
  • 9Teichroeb J H,Forrest J A,Ngai V. Imaging protein deposits on contact lens materials[J].Optometry and Vision Science,2008,(12):1151-1164.doi:10.1097/OPX.0b013e31818e8ad6.
  • 10Cho P,Ng V. Clinical performances of two disposable soft contact lenses of different materials on Hong Kong-Chinese[J].Contact Lens & Anterior Eye,2000,(02):53-60.doi:10.1016/S1367-0484(00)80026-0.