基于FPGA的高强度聚焦超声相位控制系统设计 被引量:1

Design of high intensity focused ultrasound phase control system based on FPGA
下载PDF
导出
摘要 多阵元相控阵换能器具有焦距可调和可实现经颅聚焦等优势,近来受到众多研究者的关注,其相位控制系统是决定多阵元相控换能器能否应用于临床的关键技术之一。在输出信号稳定的同时提高高强度聚焦超声相位控制系统精度是设计过程中的重点与难点。本文基于现场可编程门阵列设计了一种高强度聚焦超声相位控制系统。实测结果表明,多通道延时分辨率为1 ns,延时误差小于1 ns,可满足多阵元高强度聚焦超声治疗相控聚焦换能器延时精度的需要。 Multi-element phased array transducer has the advantages of adjustable focal length and transcranial focusing,and has recently attracted the attention of many researchers.The phase control system of the multi-element phased transducer is one of the key technologies for determining whether the multi-element phased phase transducer can be applied to the clinic.Increasing the accuracy of the high intensity focused ultrasound(HIFU)phase control system while the output signal is stable is a key and difficult point in the design process.A HIFU phase control system is designed based on field-programmable gate array(FPGA).The measured results show that the multi-channel delay resolution is 1 ns and the delay error is less than 1 ns,which can meet the needs of multi-element HIFU for phased-focus transducers.
作者 张浩 张艳秋 赵梦娟 菅喜岐 ZHANG Hao;ZHANG Yanqiu;ZHAO Mengjuan;JIAN Xiqi(Department of Biomedical Engineering and Technology,Tianjin Medical University,Tianjin 300070,China)
出处 《应用声学》 CSCD 北大核心 2018年第6期855-860,共6页 Journal of Applied Acoustics
基金 天津市自然科学基金重点项目(16JCZDJC32200) 天津医科大学科技成果转化项目
关键词 高强度聚焦超声 相控阵换能器 相位控制系统 延时分辨率 现场可编程门阵列 High intensity focused ultrasound Phased-array focusing transducer Phase control system,Delay resolution Field-programmable gate array
  • 相关文献

参考文献5

二级参考文献40

共引文献82

同被引文献15

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部