Beckman Coulter AU5800全自动生化分析仪检测血清淀粉样蛋白A的性能验证分析 被引量:5

Performance study of Beckman Coulter AU5800 automatic biochemical analyzer for detecting serum amyloid A
下载PDF
导出
摘要 目的对血清淀粉样蛋白A(SAA)试剂盒在Beckman Coulter AU5800全自动生化分析仪(以下简称"AU5800全自动生化分析仪")上的检测性能进行评价。方法通过实验计算AU5800全自动生化分析仪检测SAA的精密度、准确度、线性范围、临床可报告范围、参考区间,与厂家说明书提供的参考值比较,评估SAA试剂在AU5800生化分析仪上的检测性能。结果 SAA的高、低水平质控品的批内精密度、批间精密度的变异系数(CV)分别为1.3%、3.3%,1.3%、3.4%,均满足厂家说明书的要求;与参考方法比较:线性回归方程Y=0.985 X+1.259 5,决定系数(r2)=0.993 9,t=-0.01,P=0.992;在SAA范围为0~202.87mg/L,r2=0.998 9,符合试剂说明书r2≥0.990的要求;检验结果的最大可报告范围可扩展至1 622.96 mg/L。结论 SAA试剂盒在AU5800生化分析仪上检测性能能满足厂家和卫生行业标准要求,且其准确度好、重复性好,结果可靠,能较好地满足临床工作。 Objective To evaluate the performance of Beckman Coulter AU5800 automatic biochemical analyzer(abbreviated as AU5800 biochemical analyzer)for detecting serum amyloid A(SAA).Methods The precision,accuracy,linearity,clinical report range and reference intervals of SAA were verified by AU5800 biochemical analyzer with instructions to evaluate SAA reagent produced by Chongqing Zhongyuan Biotechnology Co,Ltd.Results The CV of high and low SAA quality control products were 1.3%,3.3%,and 1.3%,3.4%respectively,which met the requirements of manufacturer′s specifications.Compared with the reference method:linear regression equation Y=0.985 X+1.259 5,r 2=0.993 9,t=-0.01,P=0.992.The linearity range was 0-202.87 mg/L.The maximum reportable range of test results can be extended to 1 622.96 mg/L.Conclusion The performance of SAA kit on AU5800 biochemical analyzer can meet the requirements of manufacturers and health industry standards,and its accuracy is good,repeatability is good,the results are reliable,can better meet the clinical work.
作者 曹修娥 栾芳 刘义庆 王勇 张炳昌 CAO Xiu′e;LUAN Fang;LIU Yiqing;WANG Yong;ZHANG Bingchang(Department of Clinical Laboratory,Shandong Provincial Hospital Affiliated to Shandong University,Jinan,Shandong 250021,China)
出处 《国际检验医学杂志》 CAS 2018年第21期2600-2603,共4页 International Journal of Laboratory Medicine
基金 国家自然科学基金青年基金资助项目(81101484) 山东省优秀中青年科学家奖励基金(BS2011SW035)
关键词 血清淀粉样蛋白A 全自动生化分析仪 性能验证 serum amyloid A automatic biochemical analyzer performance verification
  • 相关文献

参考文献5

二级参考文献53

  • 1朱华,王娟,林燕,张芃,刘家民.血清淀粉样蛋白A的研究进展[J].心血管病学进展,2007,28(1):113-115. 被引量:17
  • 2DAN-HUI LIU,XIAO-MIN WANG,LI-JUAN ZHANG,SONG-WEI DAI,LI-YUN LIU,JI-FU LIU,SHAN-SHAN WU,SHUAN-YING YANG,SAM FU,XUE-YUAN XIAO,DA-CHENG HE.Serum Amyloid A Protein: A Potential Biomarker Correlated With Clinical Stage of Lung Cancer[J].Biomedical and Environmental Sciences,2007,20(1):33-40. 被引量:12
  • 3戴嵩玮,王小敏,刘丽云,刘吉福,武珊珊,黄凌云,肖雪媛,何大澄.一个在肺癌血清中高表达的标志分子SAA的发现及鉴定[J].中国科学(C辑),2007,37(2):129-134. 被引量:25
  • 4Uhlar CM,Whitehead AS. Serum amyloid A,the majorvertebrate acute-phase reactant [ J]. Eur J Biochem,1999, 265(2) :501-523.
  • 5Coetzee GA, Strachan AF, van der Westhuyzen DR,et al. Serum amyloid A-containing human high densi-ty lipoprotein 3. Density, size and apolipoproteincomposition [ J ]. J Biol Chem,1986, 261 (21 ):9644-9651.
  • 6De Beer MC, Webb NR, Wroblewski JM, et al. Im-pact of serum amyloid A on high density lipoproteincomposition and levels [ J]. J Lipid Res, 2010,51(11) : 3117-3125.
  • 7De Beer MC, Yuan T,Kindy MS, et al. Characteri-zation of constitutive human serum amyloid A protein(SAA4 ) s an apolipoprotein [ J ]. J Lipid Res,1995,36(3) :526-534.
  • 8Urieli-Shoval S, Linke RP, Matzner Y. Expressionand function of serum amyloid A, a major acute-phase protein,in normal and disease states [ J ].Curr Opin Hematol,2000 , 7(1) :64-69.
  • 9Hua S, Song C, Geczy CL, et al. A role for acute-phase serum amyloid A and high-density lipoproteinin oxidative stress,endothelial dysfunction and athero-sclerosis [J]. Redox Rep, 2009,14(5) : 187-196.
  • 10Sandri S , Hatanaka E , Franco AG, et al. Serumamyloid A induces CCL20 secretion in mononuclearcells through MAPK (p38 and ERK1/2) signalingpathways [J]. Immunol Lett, 2008,121 (1) : 22-26.

共引文献205

同被引文献25

引证文献5

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部