摘要
按照《环境空气铅的测定石墨炉原子吸收分光光度法》(HJ539-2015)对方法检出限、精密度和准确度进行分析,结果表明,方法的检出限为0. 008μg/m3,精密度与准确度均符合方法要求,从而证明了该方法在监测分析工作中的适用性与有效性。
The detection limit,precision and accuracy of the method are analysedby graphite furnace atomic absorption spectrophotometry for the determination of lead in ambient air(HJ539-2015).The results show that the detection limit of the method is 0.008μg/m^3,and the precision and accuracy are in accordance with the method requirements.It proved that the new method is applicable and effective in the monitoring and analysis of work.
作者
杨惠
Yang Hui(Huludao Environmental Monitoring Central Station,Liaoning Huludao 125000)
出处
《江西化工》
2018年第6期174-175,共2页
Jiangxi Chemical Industry
关键词
石墨炉原子吸收分光光度法
环境空气
铅
方法验证
graphite furnace atomic absorption spectrophotometry
ambient air
lead
method validation