期刊文献+

气相分子吸收光谱法应用于废水中的氨氮测定 被引量:4

Determination of Ammonia Nitrogen in Wastewater by Gas-phase Molecular Absorption Spectrometry
下载PDF
导出
摘要 目的:采用气相分子吸收光谱法对废水中的氨氮进行测定。方法:完成软件的准备工作并清洗完毕后,将待测样品放入仪器中,待测。结果:废水处理装置进口、出口浓度分别为1.1760mg/L、0.1806mg/L,回收率分别为99.54%、99.58%,标准偏差分别为0.4480%、0.6740%。结论:该实验方法快捷、简便、具有良好的准确度和精密度。相对于纳氏试剂法而言,具有更高的自动化程度与更快的测定速度等优势。 Objective:To determine ammonia nitrogen in wastewater by gas phase molecular absorption spectrometry.Methods:After completing the software preparation and cleaning,put the sample into the instrument for testing.Results:The concentration at the inlet and outlet of the wastewater treatment plant were 1.1760mg/L and 0.1806mg/L respectively,the recovery rates were 99.54%and 99.58%respectively,and the standard deviations were 0.4480%and 0.6740%respectively.Conclusion:The experimental method is fast,simple,and has good accuracy and precision.Compared with Nessler's reagent method,it has the advantages of higher automation and faster determination speed.
作者 李福君 Li Fujun(Tongliao Vocational College,Inner Mongolia,028000)
机构地区 通辽职业学院
出处 《当代化工研究》 2018年第12期111-112,共2页 Modern Chemical Research
关键词 气相分子吸收光谱法 废水 氨氮 测定 gas-phase molecular absorption spectrometry waste water ammonia nitrogen determination