基于电磁阀的小流量换向器修正方法研究

Research on Correction Method of Small Flow Diverter Based on Solenoid Valve
下载PDF
导出
摘要 为了研究二位三通电磁阀在微小流量下的换向性能,研究中对某型电磁阀换向时间差开展了测试试验。试验结果表明电磁阀换向时间差为负值,且时间差随流量减小其绝对值增大。在高流量段电磁阀具有良好换向性能;而在低流量范围内,通过采用双衰减函数进行修正,可有效降低换向时间差引入的不确定度。研究表明:将电磁阀作为换向器应用于小流量液体流量标准装置具有较高可行性。 In order to study the performance of the two-position three-way solenoid valve as a diverter under a small flow rate, a test was conducted to investigate the diverting time difference of a certain type of solenoid valve. The test result shows that the diverting time difference of the solenoid valve is negative,and the difference increases as the flow rate decreases. The solenoid valve has good performance in high flow range. In the low flow range,the correction method by using a double exponential decay function, the uncertainty introducend by the diverting time difference can beeffectively reduce. The study shows that it is highly feasible to use a solenoid valve as a diverter for the standard facility for small liquid flow.
作者 张永胜 于华伟 张毅治 ZHANG Yongsheng;YU Huawei;ZHANG Yizhi(Changcheng Institute of Metrology & Measurement,Beijing 100095,China)
出处 《计测技术》 2018年第6期37-40,共4页 Metrology & Measurement Technology
关键词 计量学 电磁阀 换向器 液体流量标准装置 metrology solenoid valve diverter standard facility for liquid flow
  • 相关文献

参考文献4

二级参考文献17

  • 1JJG164-2000.液体流量标准装置检定规程[S].[S].,2000..
  • 2Kawakita K, Pereira M T. A new laboratory for calibration and testing of oil meters and measurement systems compo- nents [ C ]//The 15^th International Flow Measurement Confer- ence. Taipei ,2010.
  • 3David C, Claude1 P. Novel water flow facility in France range extension to low flow rates ( 10000 ml/h down to 1 ml/h) [ C ]//The 15^th International Flow Measurement Conference. Taipei ,2010.
  • 4Ciftci V, Akselli B, Ergen E, et al. Introducing UME divert- ers' and weighting systems for water and hydrocarbon flow measurement[ C]//The 12^th International Flow Measurement Conference. Guilin ,2004.
  • 5Doihara R, Shimada T, Terao Y, et al. Development of a di- verting system employing a rotating double wing method [ C]//The 12^th International Flow Measurement Conference. Gui.lin,2004.
  • 6Engel, Rainer. Modeling the Uncertainty in Liquid Flowmeter Cali- bration and Application-Requirement and Their Technical Realiza- tion for PTB's National Water Flow Standard[ C ]//SENSOR Con- ference 2007 Proceedings Ⅱ.
  • 7Doihara R, Shimada T, Terao Y, et al. Development of Weighing Tank System Employing Rotating Double Wing Diverter[ J]. Flow Measurement and Instrumentation, 2006,17 : 141 - 152.
  • 8Gowda V,Yeh T T, Espina P I, et al. The New NIST Water Flow Calibration Facility[ C ]//Proe of 11 th FLMEKO 2003.
  • 9Kar-Hooi Cheong, Takashi Shimada, Ryouji Doihara. A New Calibration Facility for Small Flow of Hydrocarbon Liquid. 15th Flow Measurement Conference (FLOMEKO). October 13 - 15, 2010 Taipei, Taiwan.
  • 10国家质量监督检验检疫总局.JJGl64-2000液体流量装置检定规程[S].北京:中国计量出版社,2000.

共引文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部