期刊文献+

搏动式血泵的电磁驱动装置设计及可行性研究 被引量:5

Design and feasibility study of electromagnetic driving device for pulsating blood pump
下载PDF
导出
摘要 为了实现搏动式血泵的驱动,设计了一种新型电磁驱动装置。基于模拟体外循环实验对装置进行相关的可行性实验研究。研究结果表明:驱动线圈结构设计过程科学合理,且驱动力规律满足血泵驱动要求;平均灌注压与驱动电流的线性拟合方程为y=33. 074x+6. 6563,R^2=0. 9784,表明该驱动装置能够按照设定的压力值输出搏动流以满足实际应用需求。装置的驱动效率为21. 83%。研究表明我们所设计的新型驱动装置能够作为搏动式血泵驱动。 In order to realize the driving of pulsating blood pump,a novel electromagnetic driving device was put forward.Based on the simulated cardiopulmonary bypass experiment,the feasibility of the device was studied.The results of the study were as follows:The design process of the driving coil structure was scientific and reasonable,and the driving force met the requirements of the blood pump driving;The linear fitting equation of average perfusion pressure and driving current was y=33.074x+6.6563,R 2=0.9784,which indicated that the driving device could output pulsatile flow according to the set pressure value to meet practical application needs.The drive efficiency of the device was 21.83%.The research shows that the new driving device designed in this paper has a great feasibility as a pulsating blood pump.
作者 张磊 葛斌 方旭晨 张少伟 魏凌轩 伍进平 ZHANG Lei;GE Bin;FANG Xucheng;ZHANG Shaowei;WEI Lingxuan;WU Jinping(University of Shanghai for Sci-Tech,Shanghai 200093,China;East City Hospital of Shanghai Yangpu District,Shanghai 200438)
出处 《生物医学工程研究》 2018年第4期470-475,共6页 Journal Of Biomedical Engineering Research
基金 上海市大学生创新计划基金项目(2016118)
关键词 搏动泵 电磁驱动 每搏输出量 平均灌注压 驱动效率 电磁力 Pulsating pump Electromagnetic drive Per stroke output Mean perfusion pressure Driving efficiency Electromagnetic forc
  • 相关文献

参考文献4

二级参考文献90

共引文献25

同被引文献54

引证文献5

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部