摘要 为了提升锥形束CT图像质量保证的速度和准确性,该文设计并实现了一种基于特征提取的自动化图像质量保证软件。该文介绍了软件如何通过Canny算法和Hough变换圆检测实现Catphan500模体的定位和各个组件中ROI的选取。针对Catphan500模体中的不同组件,该文分别介绍了软件中HU准确性、空间线性度、HU均匀性、空间线性度的计算方法。最后,该文通过实验验证了软件的可行性。 In order to improve the speed and accuracy of cone beam CT image quality assurance, an automatic image quality assurance software based on feature extraction is designed and implemented. This paper introduces how the software can realize the positioning of Catphan500 phantoms and the selection of ROI in each module through Canny algorithm and Hough transform circle detection. For the different modules in the Catphan500 phantom, this paper describes the calculation methods of HU accuracy, spatial linearity, HU uniformity and spatial linearity in the software. Finally, the article verifies the feasibility of the software through experiments.
作者 汪涛 冯汉升 李实 杨洋 WANG Tao;FENG Hansheng;LI Shi;YANG Yang(Institute of Plasma Physics, Chinese Academy of Sciences, Hefei, 230031;University of Science and Technology of China, Hefei, 230026)
出处 《中国医疗器械杂志》 2019年第1期25-28,共4页 Chinese Journal of Medical Instrumentation
基金 中科院合肥物质科学研究院"十三五"规划重点支持项目(KP-2017-24)
关键词 图像引导放疗 锥形束CT 质量保证 自动化 Image Guided Radiotherapy(IGRT) Cone Beam Computed Tomography(CBCT) quality assurance automated
  • 相关文献

参考文献3

二级参考文献31

  • 1黄岗,严汉民.点扩散函数测量CT机MTF的应用研究[J].医疗设备信息,2004,19(8):10-12. 被引量:5
  • 2胡曙光,昌仁民,冯小刚,康立丽,林意群,张雪林.CT的MTF测量和评价[J].中国医学物理学杂志,1997,14(1):30-32. 被引量:7
  • 3Yu Lifbng,Thomas JV,Michael RB,et al.Dose and image quality evaluation of a dedicated cone-beam CT system for high contrast neurologic applications[J].AJ R,2010,194:W 193- W201.
  • 4余晓锷,卢广文.CT设备原理和质量保证[M].北京:科学出版社.2005:30-32.
  • 5Vedantham S,Karellas A.Modeling the performancecharacteristics of computed radiography (CR) systems[J].lEEE Trans Med Imaging,2010,29:790-806.
  • 6Pouliot J;Bani-Hashemi A;Chen J.放射治疗用的低剂量MV级锥束CT[J]国际放射肿瘤物理生物杂志,2005552-560.
  • 7Jaffray DA;Drake DG;Moreau M.一种影像学和层析成像系统集成到一个医疗直线加速器对骨骼和软组织的目标的定位[J]国际放射肿瘤物理生物杂志,1999773-789.
  • 8Sharpe MB;Morseley DJ;Purdie TG.带锥形束CT的直线加速器机械校准的稳定性[J]医学物理杂志,2006136-144.
  • 9Letoumeau D;Wong JW;Oldham M.锥束CT引导放射治疗:技术应用[J]肿瘤放射治疗杂志,2005(03):279-286.
  • 10Chandra A;Dong L;Huang E.超声引导的前列腺日常定位经验[J]国际放射肿瘤物理生物杂志,2003(02):436-447.