一种热释光剂量计的性能检测分析 被引量:1

Performance Analysis of A Thermoluminescent Dosimeter
下载PDF
导出
摘要 目的通过系统检测一种热释光剂量测量系统的性能指标,全面评价该系统的可靠性。方法参照JJG 593-2006、GBZ207-2016和GB/T 10264-2014,对热释光剂量测量系统的批均匀性、稳定性、入射角响应、线性、能量响应、量值检验、残余剂量、探测限等8个指标进行了检测,并将检测结果与相关标准要求比对。结果各项性能指标检测结果满足国家及行业规定的指标限值。结论该热释光剂量测量系统符合相关标准要求,可以用于人与环境辐射剂量的测量工作。 Objective To evaluate the reliability of a thermoluminescence dosimetry system by testing the key performance indexes. Methods Under the criteria of JJG593-2006, GBZ207-2016, and GB/T10264-2014, eight indexes of the thermoluminescence dosimetry system were examined, including the batch homogeneity, stability, angle response, linearity, energy response, quantity inspection, residual dose, and the minimal detection limit. Moreover, the results were compared with the related standards. Results The results of various performance indexes met the criteria of the national and industrial standards. Conclusion The thermoluminescence dosimetry system meets the technical requirements of relevant standards, and can be used for personal and environmental dose monitoring.
作者 赵进沛 王琳杰 周超 张建 江其生 李秀芹 ZHAO Jinpei;WANG Linjie;ZHOU Chao;ZHANG Jian;JIANG Qisheng;LI Xiuqin(Department of Health Protection,Beijing Military Center for Disease Control and Prevention,Beijing 100042,China;Military Preventive Medical College,Army Medical University,Chongqing 400038,China;Beijing JunYiLianRui Institute of Nuclear Measurement Technology,Beijing 100850,China;Center Lab,General Hospital of Rocket Troops,Beijing 100088,China)
出处 《中国医疗设备》 2019年第3期37-40,共4页 China Medical Devices
基金 全军立项课题(AEP17J001)
关键词 热释光 剂量检测 性能评价 个人剂量 thermoluminescence dosimetry performance evaluation personnel dosimetry
  • 相关文献

参考文献9

二级参考文献45

共引文献109

同被引文献1

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部