期刊文献+

离子色谱-安培检测法检测饮用水中氰化物的方法分析 被引量:4

Method Analysis for Determination of Cyanide in Drinking Water by IC-PAD
下载PDF
导出
摘要 利用离子色谱-安培检测器建立一种快速测定饮用水中痕量氰化物的方法.该方法最低检出限可达0.17μg/L(信噪比S/N=3),最低定量限为0.56μg/L(信噪比S/N=10),在0~150μg/L的浓度范围内线性良好,不易受饮用水中常规阴离子、硫化物及硫氰化物的影响,实际样品的测试结果与气相色谱国标法的检测结果相一致.实验表明,该方法重复性好,检测限低,精密度高,准确性高,重复性好,检测时不易受其他离子影响,适宜大批量、快速测定饮用水中的痕量氰化物含量. A method for rapid determination of trace cyanide in drinking water is established using an ion chromatography-amperometric detector.The minimum detection limit of this method is 0.17 μg/L (signal-to-noise ratio S/N = 3),the minimum limit of quantification is 0.56 μg/L (signal-to-noise ratio S/N = 10),and has a good linearity over a concentration range of 0-150 μg/L,and it is not easy to be affected by the conventional anions,sulfides and thiocyanides in drinking water.The test results align with the results of the gas chromatographic GB method when testing the actual samples.Experiments show that the method has good repeatability,low detection limit,high precision and high accuracy.It is not easily affected by other ions during the detection.It is suitable for the determination of cyanide content in drinking water.
作者 傅珠婷 石健 蔡萌 李春建 FU Zhuting;SHI Jian;CAI Meng;LI Chunjian(Analysis and Testing Center,Nantong University,Nantong 226019,China)
出处 《南通大学学报(自然科学版)》 CAS 2018年第4期61-65,共5页 Journal of Nantong University(Natural Science Edition) 
关键词 离子色谱法 安培检测法 饮用水 氰化物 ion chromatography(IC) pulsed amperometric detection(PAD) drinking water cyanide
  • 相关文献

参考文献11

二级参考文献58

共引文献158

同被引文献31

引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部