摘要 目的评价无创血流动力学监测系统(Bioz.com)用于心脏手术患者围手术期心功能评估的准确性及可靠性。方法选取在2017年12月~2018年6月于承德医学院附属医院行择期心脏手术患者35例,包含瓣膜性疾病23例,冠心病12例。选取"肺动脉漂浮导管"及"超声心动图"作为比较指标,严格按照Bioz.com系统操作规范进行操作,对心排出量(CO)、每搏输出量(SV)、心指数(CI)进行监测,分别记录手术前后患者血流动力学变化情况,并进行相关统计学分析。结果心脏手术前后,采用Bioz.com系统及超声心动图两种方法对患者血流动力学参数CO、SV、CI进行监测并比较,差异均无统计学意义(P>0.05)。心脏手术中,采用Bioz.com系统及漂浮导管两种方法对患者血流动力学参数CO、SV、CI进行监测并比较,差异均无统计学意义(P>0.05)。结论在心脏手术中Bioz.com系统与肺动脉漂浮导管,Bioz.com系统与超声心动图监测的精确度与准确度具有一致性,是一种较为可行的心功能监测方法。 Objectives To evaluate the accuracy and reliability of the non-invasive hemodynamic monitoring system (Bioz.com) for perioperative cardiac function assessment in patients undergoing cardiac surgery. Methods From December 2017 to June 2018, 35 patients underwent elective cardiac surgery at the Affiliated Hospital of Chengde Medical College, including 23 patients with valvular disease and 12 patients with coronary heart disease. The "pulmonary artery floating catheter" and "echocardiogram" were selected as comparison indicators, and the operation was carried out in strict accordance with the Bioz.com system operation specification. The cardiac output (CO), stroke volume (SV), and heart index (CI) were monitored. The hemodynamic changes of patients before and after surgery were recorded and statistical analysis was performed. Results Before and after cardiac surgery, the hemodynamic parameters CO, SV and CI were monitored and compared by Bioz.com system and echocardiography,the differences were not statistically significant (P>0.05). In the cardiac surgery, the hemodynamic parameters CO, SV and CI were monitored and compared by Bioz.com system and floating catheter,the difference was not statistically significant (P>0.05). Conclusion In the cardiac surgery, the Bioz.com system and the pulmonary artery floating catheter, Bioz.com system and echocardiography monitoring are consistent with the accuracy and accuracy, which is a feasible method for heart function monitoring.
作者 赵悦 董婕 韩晨阳 周国明 ZHAO Yue;DONG Jie;HAN Chen-yang;ZHOU Guo-ming(Department of Pain,the Affiliated Hospital of Chengde Medical College,Chengde 067000,Hebei,China;Science Press,Beijing 100000,China;Chengde Medical College,Chengde 067000,Hebei,China)
出处 《医学信息》 2019年第5期108-110,共3页 Journal of Medical Information
关键词 心脏手术 Bioz.com 心排血量 血流动力学 Cardiac surgery Bioz.com Cardiac output Hemodynamics
  • 相关文献

参考文献2

二级参考文献20

  • 1van Lelyvele-Haas LE,van Zanten AR,Bonn GF, et al.Clinical validation of the non-invasive cardiac outputmonitor USC0M-1A in critically ill patients [J]. Eur JAnaesthesiol,2008,25(11):917-924.
  • 2Tan HL,Pinder M,Parsons R,et al. Clinical evaluationof USC0M ultrosonic cardiac output monitor in cardiacsurgical patients in intensive care unit[J]. Br J Anaesth,2005,94(3):287-291.
  • 3Wong LS, Yong BH,Young KK, et al. Comparison ofUSC0M ultrasonic cardiac output monitor with pulmo-mary artery catheter thermodilution in patients undergoingliver transplantation [J]. Liver Transpl,2008,14 (7):1038-1043.
  • 4Knobloch K,Huhfich V, Rohmann P,et al. Non-inva-sive determination of cardiac output by continuous waveDoppler in air rescue service[J]. Anasthesiol IntensivmedNoffallmed Schmerzther,2005,40(12):750-755.
  • 5Chand R,Mehta Y,Trehan N. Cardiac output estimationwith a new Doppler device after off-pump coronary arterybypass surgery[J]. J Cardiothorae Vasc Anesth,2006,20(3):315-319.
  • 6Brierley J,Thiruchelvam T,Peters MJ. Hemodynamics ofearly pediatric fluid-resistant septic shock using non-in-vasive cardiac output (USC0M) :distinct profiles of CVCinfection and community acquired sepsis [J]. Crit CareMed, 2006,33(12):7-11.
  • 7Siu CW,Tse HF,Lee K,et al. Cardiac resynchroniza-tion therapy optimization by ultrasonic cardiac outputmonitoring ( USC0M) device [J]. Pacing Clin Electrophys-iol, 2007,30(1):50-55.
  • 8Dey I, Sprivulis P. Emergency physicians can reliablyassess emergency department patient cardiac output usingthe USC0M continuous wave Doppler cardiac outputmonitor[J]. Emerg Med Australas,2005,17(3): 193-199.
  • 9Heerman WJ,Doyle T,Churchwell KB,et al. Accuracyof non-invasive cardiac output monitoring (USC0M) [J].Crit Care Med,2006,34(12):61.
  • 10余珍珠,刘晓红.应用无创心排血量监测技术监测窒息新生儿心排血量的研究[J].中国小儿急救医学,2008,15(2):173-174. 被引量:5