对照CG-5数字重力仪倾斜传感器检验调整步骤,解析重力仪倾斜传感器零位补偿和灵敏度的检验调整原理,剖析倾斜传感器检验调整机理,通过实验对检验调整解析原理进行验证。结果表明,本文解析的检验调整原理正确,同时提出重力测量时倾斜传感器检验调整限差制定依据。该解析方法对于了解仪器原理、优化仪器应用及仪器的二次开发利用有参考作用。
The paper analyses the principle of tilt sensor’s zero compensation and sensitivity adjustment according to adjusting procedures of the CG-5 gravimeter tilt sensor parameters.An experiment is performed to verify the principle.The results show that the principle analyzed in this paper is correct.Then,a basis for adjusting tolerance of the tilt sensor parameters is proposed.The principle analysis will be helpful for mastering the CG-5 gravimeter.Furthermore,it is useful for optimizing the instrument and secondary development.