摘要
本文主要对HORIBA气体在线监测系统ENDA系列产品在使用过程中的常见故障进行了总结,主要有NOx数据异常、氨逃逸异常、氧量异常三大类,对造成以上问题的主要原因进行了分析,并从取样管路系统的泄漏、堵塞等方面对相应的处理方法进行了介绍。
This paper summarizes the common faults of ENDA series products of HORIBA gas monitoring system in use, mainly including anomaly of NOX data, anomaly of ammonia escape and anomaly of oxygen. The main causes of the above problems are analyzed, and the corresponding treatment methods are introduced from leakage and blockage of sampling pipeline system.
作者
郑长乐
孙杰
初银晓
Zheng Chang-le;Sun Jie;Chu Yin-xiao
出处
《电力系统装备》
2019年第3期121-122,共2页
Electric Power System Equipment