摘要
介绍了中国散裂中子源(CSNS)一期4 K低温恒温器的结构、性能测试及样品温度测试结果。测试结果显示样品座测点温度与实际样品温度存在温差,所以必须对样品温度进行校正。通过对样品区域进行热工分析,设计了可以密封氦气的样品盒,实现了测点温度与样品温度的一致。
This paper mainly introduces the structure,test results of the 4K closed cycle cryostat.The test results show that there is a difference between the sample holder temperature and the actual sample temperature,so the sample temperature must be corrected.Based on the thermal analysis of the sample area,a sample box which can seal helium is designed,and the measured temperature shows good agreement with the temperature of the sample.