麻醉机内呼吸回路消毒及灭菌的指导建议 被引量:17

Recommendations for disinfection and sterilization of breathing circuit in anesthesia machine
原文传递
导出
摘要 麻醉机易被微生物污染,且其内呼吸回路需反复使用,如消毒不当,则可能造成患者之间的交叉感染。为了规范麻醉机内呼吸回路的消毒及灭菌的方法,中国心胸血管麻醉学会围术期感染控制分会成立了专家工作组,在多次讨论和修改后,制订出首版"麻醉机内呼吸回路消毒及灭菌的指导建议",欢迎读者提出意见和建议。 The anesthesia machine is easily contaminated by microorganisms, the internal breathing circuit is used repeatedly, and improper disinfection may cause cross-infection.To standardize the method of disinfection and sterilization of the internal breathing circuit in the anesthesia machines, Perioperative Infection Control Society of Chinese Society of Cardiothoracic and Vascular Anesthesiology established an expert task force.The first edition of the recommendations was developed after numerous discussions and revisions.Opinions and suggestions are warmly welcome.
作者 中国心胸血管麻醉学会围术期感染控制分会“麻醉机内呼吸回路消毒及灭菌”工作组 王古岩 曹英浩 程颢 池萍 丁超 杜英杰 范婷 关雷 高金贵 顾秀武 韩辉 李军 吕雯斐 柳小丽 孙永兴 王怀江 袁晓宁 郑晖 Perioperative Infection Control Society of Chinese Society of Cardiothoracic and Vascular Anesthesiology Task Force on Disinfection and Sterilization of Breathing Circuit in Anesthesia Machine;Wang Guyan(Department of Anesthesiology,Beijing Tongren Hospital,Capital Medical University Beijing 100730,China)
出处 《中华麻醉学杂志》 CAS CSCD 北大核心 2018年第12期1417-1420,共4页 Chinese Journal of Anesthesiology
关键词 消毒 灭菌 麻醉机 呼吸回路 Disinfection Sterilization Anesthesia machine Breathing circuit
  • 相关文献

参考文献7

二级参考文献40

  • 1应瑛.人工鼻在气管切开患者气道湿化中的应用[J].浙江临床医学,2007,9(4):533-534. 被引量:7
  • 2贾文祥.医学微生物学[M].北京:人民卫生出版社,2010:376-377.
  • 3DU MOULIN G C, SAUBERMAN A J. The anesthesia machine and circle system are not likely to be source of bacterial con- tamination[ J ]. Anesthwsiology, 1977,47 (4) :353-358.
  • 4HARTMANN D, JUNG M, NEYBERT T R, et al. Microbiologi- cal risk of anaesthetic breathing circuits after extend use [ J ]. Acta Anaethesiol Scand ,2008,52(3 ) :432-436.
  • 5LOFTUS R W, MUFFLY M K, KOFF M D, et al. Hand con- tamination of anesthesia providers is an important risk factor for intraoperative bacterial transmission [ J ]. Anesth Analg, 2011,112:98-105.
  • 6CANTRELL D, SHAMRIZ O, COHEN M J, et al. Hand hy- giene conpliance by physicians:marked heterogeneity due to local culture? [ J]. Am J Infect Control,2009,37:301-305.
  • 7HELLING T S, KRANTZ S, BERTRAM K, et al. The value of clinical judgment in the diagnosis of the diagnosis of nosocomi- al pneumonia[ J]. Am J Surg, 1996,171 (6) :570-575.
  • 8Kohlenberg A, Schwab F, Behnke M, et al. Pneumonia asso- ciated with invasive and noninvasive ventilation:an analysis of the German nosocomial infection surveillance system database [J]. Intensive care medicine, 2010,36 (6) : 971-978.
  • 9Gutelius B,Perz JF,Parker MM,et al. Multiple dusters oi hepa- titis virus infections associated with anesthesia for outpatient en- doscopy procedures[J]. Gastroenterology,2010,139(1) : 163-170.
  • 10叶英妩,王毓三,申子瑜.全国临床检验操作规程[S].南京:东南大学出版社,2006:78.

共引文献33

同被引文献85

引证文献17

二级引证文献74

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部