期刊文献+

新型多功能颅脑外引流固定仪的设计及其性能研究 被引量:2

Design and performance of a new multifunctional fixer for external brain drainage
下载PDF
导出
摘要 设计一种新型多功能颅脑外引流固定仪,其包含引流固定支架和引流液体测量模块。引流固定支架由支架、标尺、挂钩等构成;引流液体测量模块利用薄膜谐振型Lamb波传感器,依据矢量测量原理,采集流体模拟信号,电气控制模块将模拟信号转换为数字信号,获取流体流量、流速测量结果。采用设计仪器展开实验研究,结果显示,仪器测量引流液体流量误差小于1%;薄膜谐振Lamb波传感器稳定性变化情况表明仪器稳定性强,适用于测量微小通道液体的流速与流量,新型多功能颅脑外引流固定仪为引流护理提供一种高性能的医疗手段。 To design a new multifunctional craniocerebral external drainage fixer.It contained two parts,one was the drainage fixation bracket and the other was the drainage liquid measurement module.Drain fixed bracket consisted of bracket,ruler,hook,etc.The drainage liquid measurement module used thin film resonant Lambwave sensors to collect fluid analog signals based on vector measurement principles.The electrical control module converted analog signals into digital signals to obtain fluid flow and flow rate measurement results.The experimental results show that the error of liquid flow is less than 1%.The stability of thin film resonance Lambwave sensor shows that the instrument has strong stability and is suitable for measuring the flow rate and flow rate of tiny channel liquids.The new multifunctional craniocerebral external drainage fixation instrument provides a high-performance medical method for drainage nursing.
作者 姜楠 龙云玲 JIANG Nan;LONG Yunling(Beijing Tiantan Hospital,Capital Medical University,Beijing 100071,China)
出处 《生物医学工程研究》 2019年第1期115-119,共5页 Journal Of Biomedical Engineering Research
基金 北京市保健科研项目(京13-01)
关键词 新型多功能 颅脑外 引流固定仪 薄膜谐振型 Lamb波传感器 流速 New multifunctional Extracranial Drainage fixator Film resonant type Lamb wave sensor Flow rate
  • 相关文献

参考文献10

二级参考文献71

  • 1阚瑞峰,刘文清,张玉钧,刘建国,董凤忠,高山虎,王敏,陈军.可调谐二极管激光吸收光谱法测量环境空气中的甲烷含量[J].物理学报,2005,54(4):1927-1930. 被引量:44
  • 2阚瑞峰,刘文清,张玉钧,刘建国,董凤忠,王敏,高山虎,陈军,王晓梅.可调谐二极管激光吸收光谱法监测环境空气中甲烷的浓度变化[J].中国激光,2005,32(9):1217-1220. 被引量:32
  • 3陈诺琦,涂梅,赵闽方,张秀芬.高血糖对糖尿病合并急性脑梗塞的影响[J].中国糖尿病杂志,1996,4(1):18-20. 被引量:73
  • 4许雪燕.模糊综合评价模型的研究及应用[D].成都:西南石油大学,2011.
  • 5ROSENTHAL G. Long-term omcomes following decompmssive craniectomy for severe traumatic brain injury-how long should we wait to evaluate results[J]. Crit Care Med, 2011, 39(11): 2575 -2576.
  • 6JONES C. Glasgow coma scale[J]. Am J Nurs, 1979, 79(9): 1551- 1553.
  • 7JENNETT B, SNOEK J, BOND MR, et al. Disability after severe head injury: observations on the use of the glasgow outcome scale[J]. J Neurol Neurosurg Psychiatry, 1981, 44(4): 285-293.
  • 8HO KM, HONEYBUL S, LIq~FON E. Delayed neurological recov-ery after decompressive craniectomy for severe nonpenetrating traumatic brain injury[J]. Crit Care Med, 2011, 39(11): 2495-2500.
  • 9HARVEY HH. Reducing traumatic brain injuries in youth sports: youth sports traumatic brain injury state laws, January 2009-De- cember 2012[J]. Am J Public Health, 2013, 103(7): 1249-1254.
  • 10MATHIEU L, BAZILE F, BARTHELEMY R, et al. Damage control orthopaedics in the context of battlefield injuries: The use of temporary external fixation on combat trauma soldiers[J]. Orthop Traumatol Surg Res, 2011, 97(8): 852-859.