期刊文献+

一种新型机械灌注泵的性能模拟

Performance simulation on a new type of machine perfusion pump
下载PDF
导出
摘要 介绍一种新型机械灌注泵的结构与工作原理,通过CFD模拟对该泵的工作过程进行了仿真。结果显示:液体流入泵时会发生扩散,液体速度从对称轴处往外逐渐减小;液体流出泵时速度分布均匀;液体流动的湍流动能很小;泵的出口质量流量和永磁体端面受力大小由永磁体的运动方式决定,当永磁体运动的加速度以分段函数变化时,出口质量流量能长时间保持稳定;当加速度以正弦函数变化时,流出液体的搏动性更显著,并且流量大小能被更准确得控制。 To introduce the structure and working principle of the new machine perfusion pump.CFD method is utilized to analyze the working process of the new pump.The result shows that the liquid diffuses when the liquid flows into the pump,and the velocity of the liquid gradually deceases outward from the axis of symmetry,the liquid has a uniform velocity distribution when the liquid flows out of the pump,and the turbulent kinetic energy of liquid flow is very small.The rate of the outlet mass flow and the force on the end face of permanent magnet is determined by the motion of the permanent magnets.When the acceleration of the motion of the permanent magnet changes as a piecewise function,the mass flow rate of the pump can remain stable for a long time.When the acceleration of the motion of the permanent magnet changes as a sine function,the pulsation of the liquid is more pronounced,and the mass flow rate can be more accurately controlled.
作者 魏凌轩 葛斌 张磊 张少伟 伍进平 方旭晨 WEI Lingxuan;GE Bin;ZHANG Lei;ZHANG Shaowei;WU Jinping;FANG Xuchen(School of Medical Instrument and Food Engineering,University of Shanghai for Science and Technology,Shanghai 200093,China;Shidong Hospital,Shanghai 200092)
出处 《生物医学工程研究》 2019年第1期120-123,共4页 Journal Of Biomedical Engineering Research
关键词 机械灌注 CFD模拟 瞬态分析 运动方式 质量流量 Machine perfusion Pump CFD simulation TransientAnalysis Operation mode Mass flow rate
  • 相关文献

参考文献2

二级参考文献46

  • 1余洪龙,解启莲.ECMO血泵研究进[J].医学信息(医学与计算机应用),2014,0(11):543-543. 被引量:2
  • 2温越琼,王华军.通电螺线管中磁场的计算机模拟[J].西华大学学报(哲学社会科学版),1999,20(4):17-19. 被引量:2
  • 3周成斌,张镜芳,章晓华,陈萍,庄建,钱忆之.离心泵在小儿体外循环中的应用[J].广东医学,2005,26(3):335-336. 被引量:2
  • 4杨天宇,龙村,李春华,胡小琴.离心泵在心血管领域中的应用——附66例临床病例总结[J].中国循环杂志,1996,11(11):678-680. 被引量:2
  • 5Ravikumar R, Jassem W, Mergental H, et al. Liver transplantation after ex vivo normothermic machine preservation: a Phase 1 (first-in- man) clinical trial[J]. Am J Transplant, 2016,16(6) :1779-1787.
  • 6Perera T, Mergental H, Stephenson B, et al. First human liver transplantation using a marginal allograft resuscitated by normothermic machine perfusion[J]. Liver Transpl, 2016,22(1 ) :120-124.
  • 7Guarrera JV, Henry SD, Samstein B, et al. Hypothermic machine preservation in human liver transplantation: the first clinical series [J]. Am J Transplant, 2010,10(2) :372-381.
  • 8Guarrera JV, Henry SD, Samstein B, et al. Hypothermic machine preservation facilitates successful transplantation of "orphan" extended criteria donor livers[ J]. Am J Transplant, 2015,15 ( 1 ) : 161-169.
  • 9Dutkowski P, Schlegel A, de Oliveira M, et al. HOPE for human liver grafts obtained from donors after cardiac death [ J ]. J Hepatol, 2014,60(4) :765-772.
  • 10Rosenthal JT, Herman JB, Taylor R J, et al. Comparison of pulsatile machine perfusion with cold storage for cadaver kidney preservation [ J ]. Transplantation, 1984,37 ( 4 ) : 425-426.